(warming unit) WARMING BLANKET, adult 3/4 reusable 92057
                    STD
                    EEMDWAUA001
                
                Valid Article
                            Account code:
                            60100
                        
                        
                            HS Code:
                            903300
                        
                        
                            Last Updated on:
                            13/08/2025, 22:03:51
                        
                    
                Former
                    Code(s):
                     -X EHOEZFR0617 EHOEWARU101 EHOEZNL0286
                
            
            
                            
                CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
            
            
                            
                            
                In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
            
            
                            
                The product is part of at least one Kit. 
A kit is a collection of products (medical and/or logistic) that are needed for a certain intervention in emergency. The choice and quantity of the articles reflects the MSF protocols for this specific situation. The use of Kits allows to start an intervention without a detailed evaluation. 
            
                    (warming unit) WARMING BLANKET, 3/4 body, reusable
Definition
Reusable device intended to help prevent a patient from cooling prior to, during, and after a surgical intervention by serving as a heat-generating cover.
The blanket must be connected to a warm air generator
Specifications
                                    Some restricted information has been hidden. Sign in
                                     to see this information
                                    
                                
                            Technical specifications
- Washable, non allergenic material
- Autoclavable
- Made from woven textiles that will not liberate fluffs or other particles:
- patient's side: polyester micro fibres (MicroPES)
- outer side: polyurethane coated polyester micro fibres
- Designed with integrated pockets for the warm air circulation
- Outlet to connect the hose of the warming unit with elastic strap
- Blanket must only be connected to warm air generator with maximum air flow 100 m3/ hour and maximum air temperature 46°C
- Reusable (can be reused at least 150 times)
- Non sterile
Dimensions
- EEMDWAUA001: adult 3/4 blanket:
- Covers 3/4 part of the body
- Size: approx. 110 x 150 cm
- Weight: approx. 350 g
Instructions for use
The light blue side of the blanket shall always be used towards the patient.
The dark blue side is the outer side.
                                    Some restricted information has been hidden. Sign in
                                     to see this information
                                    
                                
                            Precautions for Use
- Monitor the patient's body temperature at regular intervals.
- Do not use on ischaemic limbs, as heat injuries may occur.
- Never obstruct the patient's free respiration with the blanket.
Maintenance
- Wash before first use and after each use in a washing machine at max 95°C.
- Wash separately the first 2 washes.
- Tumble dryer at max. 120°C.
- Autoclave at max. 134°C / 2 bar.
- Iron at low temperature (synthetic material).
- Do not dry-clean and do not chlorine bleach.
MSF requirements
The blanket shall be used for the optimization of patient's body temperature during surgical interventions or in the emergency room. It is reusable, easy to clean, soft and comfortable, easy to form around the patient.
                                    Some restricted information has been hidden. Sign in
                                     to see this information
                                    
                                
                            