(Determine rapid test) TUBE, CAPILLARY, EDTA 7D2222/7D2227

STD ELABTUCA1E-

Valid Article

Former Code(s): ELAETUCA1E-
Single use
A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
W0501010203 - Blood collection, capillary tubes
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
Thermosensitive codes are defined for storage and transportation temperature requirements of the products.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

(Determine rapid test) CAPILLARY TUBE

Definition

Very fine glass tube with EDTA anticoagulant, intended for sampling blood by capillarity.

Used with the HIV-1/2 Determine® rapid test.

Specifications

Some restricted information has been hidden. Sign in to see this information

Material

Glass

Image

Technical specifications

  • Capacity: ± 50 µl if filled between the two marks

NB: As the tube can be used on both sides, it is marked twice. To obtain the necessary volume of 50 µl, it has to be filled until the second mark.

  • Length: 75 mm
  • External diameter: 1.8 mm
  • Internal diameter: 1.4 mm
  • Total volume: 83 µl
  • Anticoagulant: EDTA, concentration EDTA 600 ± 75 ug/capillary = 2.3 mg/ml

Packaging & Labelling

Instructions for use

To be used in combination with the specific buffer solution Determine Chase Buffer ref.7D2247 (see related articles).

Some restricted information has been hidden. Sign in to see this information

Storage

  • Store between 15 and 30°C. Do not refrigerate.
  • Shelf life: 24 months

(Cf Introduction: Thermosensitive products)

Waste management

Immediately after use, dispose of the tube into a sharps container (see related articles below).

MSF requirements

Specific capillary tube for collecting a blood sample at the finger tip for the HIV Determine rapid tests.