CD4 FLOW CYTOMETER (Partec Cyflow II) absol & %, 3 detectors

STD ELAECD4E2--

Valid Article

Former Code(s): ELAECYFC2--
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
W0202040192 - Cytoflowmeters - ivd medical device software
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
The order of this product needs to be justified and is only acceptable under certain conditions.
Thermosensitive codes are defined for storage and transportation temperature requirements of the products.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

CD4 FLOW CYTOMETER (Partec Cyflow II)

Please contact the laboratory advisor in your OC for advise regarding the most adapted solution for CD4 count.

Definition

Partec CyFlow II Counter is a flow cytometer for the quantification of CD4+ T cells and the measurement of the CD4% of lymphocytes in blood.

The CD4+T cell concentration is an indicator for the initiation and follow-up of treatment for people living with HIV, in conjuction with other tests and clinical findings.

Specifications

Remark: Specific reagents are compatible with CyFlow Counter CY-S-3021 (1 detector) and SL-3 (3 detectors)

Some restricted information has been hidden. Sign in to see this information

Technical specifications

  • Instrument equipped with an air-cooled laser, 3 detectors (right angle scatter = Side Scatter and 2 colours), an integrated computer/printer. Cell populations are gated manually.
    • Absolute CD4 : manual gating on mono-parametric representation with 1 fluorescent antibody (CD4)
    • CD4 % : manual gating on bi-parametric representation with 2 fluorescent antibodies (CD45 and CD4)
  • Absolute counts (cell concentration/μl) are obtained by true volumetric system
  • Sample: 20µl venous whole blood in EDTA blood tube
  • No hemolysis
  • Throughput: 100 samples/day including sample preparation

Dimensions

  • Size : 325 x 330 x 265 mm
  • Weight : 11.5 kg
  • Power : 60 W

Supplied with the Article

  • Software Cyview
  • 1 UPS
  • 3 automatic pipettes
  • 1 polycarbonate tube holder for 3.5ml tubes
  • 1 sheath fluid bottle and 1 waste bottle
  • 1 plastic cover
  • 1 set of spare tubings
  • 5 joints
  • 5 spare fuses
  • Sheath fluid filters
  • Metal thread
  • 1 set of 7 screwdrivers

To be Ordered Separately

  • Blood collection materials
  • CD4 absolute count kit or CD4% + absolute count kit
  • Count check beads
  • Cleaning solution
  • Sheat fluid
  • Decontamination solution
  • Hypochlorite solution
  • Sample tubes

Instructions for use

Some restricted information has been hidden. Sign in to see this information

Precautions for Use

  • Air conditioning is necessary.
  • Dust-free environment.
  • The installation should be performed by an authorised Partec agent.
  • Secure all biomedical electrical devices in the laboratory. Contact your biomedical referent to define the best option. Failing that, use a UPS double conversion for the electrical protection of the apparatus.

Maintenance

A maintenance contract must be subscribed.

Waste management

Liquid waste must be decontaminated with bleach before discarding it.