(spectrophotometer) KIT, CRP standard 2 x 1ml 11341
                    STD
                    ELAESPET11341
                
                Valid Article
                            Account code:
                            60200
                        
                        
                            HS Code:
                            901890
                        
                        
                            Last Updated on:
                            29/06/2025, 22:39:35
                        
                    
                Former
                    Code(s):
                     DDGTSPEP111
                
            
            
                            
                CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
            
            
                                    18
                            
                Shelf life = length of time a product can remain active and effective. 
Short shelf life = shelf life = < 24 months
            
                                    W01021109 - C-reactive protein
                            
                European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
            
                            
                            
                Thermosensitive codes are defined for storage and transportation temperature requirements of the products.
            
                    (spectrophotometer) KIT, CRP
Definition
Kit of reagents to determine the dosage of C-reactive protein (CRP) in a blood sample (serum, heparin or EDTA plasma) with the use of a spectrophotometer.
Specifications
                                    Some restricted information has been hidden. Sign in
                                     to see this information
                                    
                                
                            Components
- CRP reagent kit, ref. 11241
- 1 x 40 ml ml CRP Buffer (BUF)
- 1 x 10 ml CRP Antiserum (AS)
- CRP Standard, ref. 11341
- 2 x 1 ml CRP standard (STD)
Technical specifications
- CRP Reagent kit for 40 measurements (controls, standard, samples)
- Immunoturbidimetric test for C-reactive protein (340 nm)
- Buffer: colourless fluid
- Antiserum: yellowish fluid
Packaging & Labelling
- CRP kit
- Kit with 1 vial of 40 ml and 1 vial of 10 ml
- CRP standard
- 2 x 1 ml CRP standard
Transport Dangerous Goods
Transport not regulated
Instructions for use
CAUTION: Only use the CRP kit with cuvettes. Using the flow cell for turbidimetric tests may clog the tubings.
                                    Some restricted information has been hidden. Sign in
                                     to see this information
                                    
                                
                            Storage
- Keep refrigerated between 2 and 8ºC
- After opening, the STD is stable for 6 weeks when sealed and stored at 2 and 8°C
- Do not freeze
- Shelf life: 18 months
- Guaranteed minimum remaining shelf life at delivery: 1/3 of total shelf life
- Not to be used beyond expiry date
Waste management
Substance with hazardous characteristics. Please contact your watsan referent for advice on proper disposal.
Detailed hazard and precautionary information can be found in the safety data sheet (SDS).
Classification EC Regulation N° 1272/2008
CRP Buffer:
- Eye Irrit. 2; H319 ; Causes serious eye irritation.
Signal Word
Warning
| H319 | Causes serious eye irritation | 
| P264 | Wash … thoroughly after handling | 
| P280 | Wear protective gloves/protective clothing/eye protection/face protection | 
| P305 + P351 + P338 | IF IN EYES: rinse cautiously with water for several minutes. If the victim wears contact lenses, remove them if possible. Keep rinsing | 
| P337 + P313 | If eye irritation persists: get medical advice/attention | 