GLOVE, EXAMINATION, latex, s.u. non sterile, large

STD SMSUGLOE1L-

Valid Article

Former Code(s): -X SMSUZRU0010 SMSUZYN1001
Single use
A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
T010201 - Latex examination / treatment gloves
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
Thermosensitive codes are defined for storage and transportation temperature requirements of the products.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

EXAMINATION GLOVE, latex

Definition

Non sterile glove used to protect both patient and staff from infections transmitted through contact with the hands.

Synonym

protective glove, procedure glove

Specifications

Quality Standards Comment

Technical specifications

  • Natural latex, non powdered
  • Ambidextrous shape (fits either hand)
  • Cuff: straight or rolled edge
  • Low residual extractable protein level
  • Non sterile, for single use
  • Dimensions: 3 sizes: small (S), medium (M) and large (L)
SizeMinimum lengthPalm width
small240 mm80 ± 10 mm
medium240 mm95 ± 10 mm
large240 mm110 ± 10 mm

Packaging & Labelling

Cardboard distribution box of 100 gloves.

The labelling must carry the following information:

  • "(product) containing natural rubber latex which may provoke allergic reactions, such as anaphylactic shock "
  • "non powdered glove"

Instructions for use

Some restricted information has been hidden. Sign in to see this information

Precautions for Use

  • Wash hands before and after use of gloves
  • To be worn only on dry hands
  • Discard defective or deteriorated gloves.
  • Remove and replace gloves which are damaged while in use
  • These gloves do not protect against physical and chemical risks which require other types of gloves
  • In case of allergy to latex, order latex-free gloves such as nitrile gloves (see related articles)

Storage

  • Store below 30°C, Protect from sunlight ‐ Protect from humidity
  • In the original packaging

MSF requirements

Non powdered glove, allowing the use of hydroalcoholic solution as hand cleanser.

Some restricted information has been hidden. Sign in to see this information