LEAD SHEET, lacquered, self-adhesive, 1mm thick, 5mx1m

STD EDIMRPSH2SA

Valid Article

Former Code(s): EDIMZFR0019 EDIMRPSH1SA EDIMRPSHH1SA
T030402 - Radiotherapy protection sheaths
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

RADIO PROTECTION LEAD SHEETING, self-adhesive

Definition

Rolled sheet lead is used to line the walls of an x-ray room to ensure safety and compliance with international radiation safety standards.

The lead sheets are coated and thus avoid direct contact with lead. The coating with a self-adhesive film facilitates the fixing.

Specifications

Material

Standard sheeting is rolled from 99.5% pure lead or more.

Technical specifications

  • Thickness sheet lead: 1.0 mm
  • Length: roll of 5 m
  • Weight of the roll: 57 kg
  • Can be easily cut with a regular cutter.

Instructions for use

To be applied to walls to all four walls and the door of an x-ray room to the height of 2 meters, with an overlap of 10 mm at all seams.

Image

MSF requirements

The lead layer is 1mm thick which is now the new standard for MSF.

For further information regarding radiation protection requirements please contact the Diagnostic Imaging Working Group (diagnostic-network@msf.org).