HAND-HELD ULTRASOUND (Lumify) + probes S4-1, C5-2, L12-4
Valid Article
HANDHELD ULTRASOUND (Philips Lumify) + S4-1, C5-2, L12-4
The technical specifications are in the process of being defined. For more information, contact the Diagnostic Imaging Working Group (diagnostic.network@amsterdam.msf.org).
Definition
The Philips Lumify is a portable handheld ultrasound device used to visualize internal body structures. Clinical uses include lung, cardiac, abdominal, pelvic, musculoskeletal, vascular (including venous access), nerve, soft tissue, and superficial imaging.
Specifications
Components
The following accessproes are included in EDIMULSE5A-
- 1 Lumify L12-4 Linear Array Transducer (Lung, Musculoskeletal, Soft Tissue, Superficial, Vascular)
- 1 Lumify C5-2 Curved Array Transducer (Abdomen, Gallbladder, Lung, OB/GYN)
- 1 Lumify S4-1 Broadband Sector Array Transducer (Abdomen, Cardiac, FAST, Lung, OB/GYN)
- 1 Samsung Galaxy Tab Active Pro4 10”, 5 year warranty
- 1 Samsung tablet cover
- Extra transducer cable
Technical specifications
- Imaging features
- 2D
- Colour Doppler
- M-mode
- advanced XRES and multivariate harmonic imaging
- SonoCT
- The images and clips stored in the compatible smartphone or tablet can be exported via WiFi in DICOM format to a PACS, to a shared disk or by email in formatpng / mpeg.
- MicroB USB transducers with replaceable cable
- L12-4 Linear array transducer
- Bandwidth: 12-4 MHz
- Field of view: 34.5 mm
- Footprint: 34 mm
- Scan depth: up to 12 cm
- Applications: Lung, MSK, Soft Tissue, Superficial, Vascular
- Centre line marker
- C5-2 Curved array transducer
- Bandwidth: 5-2 MHz
- Field of view: 67.3°
- Footprint: 50 mm
- Scan depth: up to 30 cm
- Applications: Abdomen, Gallbladder, Lung, OB/GYN
- Centre line marker
- S4-1 Broadband sector array transducer
- Bandwidth: 4-1 MHz
- Field of view 90°
- Footprint: 20.2 mm
- Scan depth: up to 30 cm
- Applications: Abdomen, Cardiac, FAST, Lung, OB/GYN
- Centre line marker
To be Ordered Separately
Recommended additional orders
MSF code | Item | Notes |
| EEMDPOCS05TEB | POWER CORD, 5m, 10A, 230V, type E+F/EUR, C13 connector,black | Recommend 1 backup depending on plug type |
| EEMDPOCS05TGB | POWER CORD, 5m, 10A, 230V, type G/UK, C13 connector, black | Recommend 1 backup depending on plug type |
EDIMULSC1CA* | Ultrasound Gel, 250 ml bot.) | Consumption: 1 bottle = 10 exams, order min 40/month |
| SDISSUQA2B-** | Disinfection – Surfanios DISINFECTANT for med. equip. For surfaces, 2 L tin + dosing pump | Diluted solution needed for daily IPC |
- This gel has a long shelf-life with no temperature requirements.
Instructions for use
Maintenance
The transducer is not autoclavable.
After each examination, the transducer must be cleaned and disinfected with a detergent-disinfectant solution for medical equipment (either 25 ml = 1 stroke of dosing pump for 5 liters of water).
(Cf Introduction: Disinfection and sterilization in the field)
Storage
- Temperature +10°C - +40°C
- Relative humidity: 15 - 95 %
- Atmospheric pressure: 700 - 1050 hPa
MSF requirements
- In contrast, the Lumify provides an ultra-portable, cheaper option, with a battery life that lasts all day.
- It can be an ideal second machine in sites that already have training
- It is also ideal with sites that need a machine kept for a specific service or group (such as in the OT for Anesthesia)