PATIENT TRANSFER MAT, 185 x 75 cm, 6 handles

STD EHOEPTRAM2-

Valid Article

CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
V080504 - Patient transfer boards and tables
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

PATIENT TRANSFER MAT with handles

Definition

A manual device used by healthcare professionals to assist in the physical movement and transfer of a person/patient (e.g., a hospitalized, a person with a disability, or a geriatric) from one position to another.

The device typically has no lifting capabilities and utilizes sliding techniques to move the person. It is typically designed using low-friction artificial fibres and may be designed as an open roll, rather like "a sleeping bag", or a flat sheet.

It is slid under the person to be moved, the person is gently pushed by the attending staff and it rolls whilst the person glides on the rotating surface.

Synonym

rescue mat

Specifications

Quality standards

EN 1865-1, 2010, +A1 2015 Patient handling equipment used in road ambulances. Part 1: General stretcher systems and patient handling equipment

Technical specifications

  • with handle loops for carrying: minimum 6 (3 per side including corners)
  • robust synthetic material: PVC, nylon, fire resistant
  • non-absorbent and impermeable to liquids.​
  • width of the tape of the handles ensures a secure and safe grip
  • radiolucent
  • lifting capacity: 150 kg minimum
  • length: 185 cm ± 10 mm
  • width: 76 cm ± 5 mm
  • dimensions wrapped: 15 x 8 x 76 cm
  • tapes width: 5 cm
  • handles length: 17 cm
  • weight: 1.5 kg

Instructions for use

Maintenance

Can be cleaned and disinfected with a detergent/disinfectant solution for surfaces / non-invasive medical equipment. Prepare use-solution in concentration required. Always read the label and product information before use. Do not mix with other products. Make sure to wet surfaces completely, use a wipe or towel, keep them wet for the whole exposure time.

MSF requirements

Intended to transfer patients from the stretcher to the bed or for transport from the rescue place to the stretcher for patients who do not have spinal injuries.

Some restricted information has been hidden. Sign in to see this information