RADIATION PROTECTION WINDOW 2.1mm lead equivalent
STD
EDIMRPWI1--
Valid Article
Account code:
60100
HS Code:
902290
Last Updated on:
07/11/2024, 20:35:49
Former
Code(s):
-X
T030403 - Radiotherapy protective screens
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
RADIATION PROTECTION WINDOW
Definition
Lead glass is used to create an observation window within a radioprotective cubicle/wall inside an x-ray room. A lead-lined protective wall provides a safe space for radiographers to work behind and the window made of lead glass allows the radiographer to monitor the patient while taking an x-ray.
Specifications
Material
Glass:
- Lead barium type
- High quality optical grade
- > 60% heavy metal oxide, including at least 55% PbO
Technical specifications
- Thickness of lead protection: min 2.1 mm of lead equivalent
- Outer dimensions vary depending on supplier: approximately 8 mm thick
Dimensions
The size of the window should be at least 30 cm x 30 cm. Exact dimensions available will vary depending on manufacturers
Instructions for use
Precautions for Use
- When installing, care should be taken that the sealing agents do not contain any acid or alkaline substances (e.g. ammonia).
- When cleaning, use only mild detergents.
Some restricted information has been hidden. Sign in
to see this information