(MC2/SEAL) 5 BLADE ELECTRODES straight ins.65mm,reus.V11CCDI

STD EEMDESUA402

Valid Article

Former Code(s): EHOEELSU402
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
Z12010980 - Electrosurgery instruments – hardware accessories
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

(ESU) ACTIVE MONOPOLAR ELECTRODE

Definition

The active electrode delivers the RF current from the generator to the surgical site. The current to tissue action at the tip of the active electrode is high electron flow over a small area. As resistance is encountered heat is produced. The current density differs for each tip based on the size and shape of the tip.

With monopolar electrosurgery, a probe electrode (pencil-like) is used to apply the electrosurgical energy to the target tissue to achieve the desired surgical effect. The current then passes through the patient to a return electrode and then back to the ESU generator to complete the circuit. Reusable monopolar electrodes have to be used with a (reusable) handpiece. The diameter of the standard shaft is 2.4 mm

Image

Specifications

Some restricted information has been hidden. Sign in to see this information

Technical specifications

monopolar electrode

  • diameter of shaft = 2.4 mm
  • blade tip
  • length: short (+/- 6.5 cm) or long (+/- 16cm)
  • reusable

handpiece

  • stainless steel
  • handle with finger switches: yellow switch = actuation of cutting current. blue switch = actuation of tissue coagulation.
  • with silicone cable (4.6 m) to connect to the electrosurgical generator
  • fits the electrodes with a shaft diameter of 2.4 mm.
  • to sterilize by steam
  • 50 uses

To be Ordered Separately

  • SSCOETCP55S: CLEANING PAD for electrosurgical tip, 5x5cm, sterile

Active electrode tips made from stainless steel have the disadvantage of eschar build up which results in increased resistance. This resistance interferes with the desired tissue effect and can become a fire hazard. These tips have to be cleaned with a scratch pad.

Instructions for use

Precautions for Use

The electrode with the proper shaft diameter must be firmly attached to the handle tip.

Active HF instruments and HF handles must not be placed on the patient in order to avoid any injuries in case of involuntary HF activation.

Maintenance

The reusable electrode-handpiece must be cleaned and sterilized by steam autoclave after each use. Recommended by the manufacturer for 50 cycles at 121 degrees for 20 minutes, after proper cleaning and drying.

(Cf Introduction: Disinfection and sterilization in the field)