CENTRIFUGE, electrical 220V (Hettich EBA 20), 8 tubes

NST ELAECENE4E-

Discontinued Article

Former Code(s): -X ELAECENT4E-
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Documents See Related

Add article to list

   

Add Kit To Wishlist

   

CENTRIFUGE, electrical 220V (Hettich EBA 20)

The model EBA 20 is stopped in 2015 because it is no longer produced and replaced by the model EBA 200 (see “replaced by” at bottom of the page).

The spare-parts for the model EBA 20 can still be ordered.

Definition

Rapid rotating apparatus, used to centrifuge biological fluids (blood, cerebrospinal fluid, urine).

Specifications

Some restricted information has been hidden. Sign in to see this information

Technical specifications

  • Plastic body, steel base
  • Model EBA 20, 8 tubes
  • Variable speed: 500 - 6000 rpm (with intervals of 100 rpm)
  • RCF (Relative Centrifugal Force): max. 3421 x g
  • 0 to 99 minutes timer
  • Fixed angle rotor for 8 tubes (15 ml) of max. 120 mm long and max. 17 mm in Ø
  • Power supply: 220 V AC / 50-60 Hz
  • Consumption: 0.28 A

Dimensions

  • Width x Depth x Heigth: 231 x 292 x 216 mm
  • Weight: approx. 4 kg

Supplied with the Article

  • 1 connecting cable
  • 2 fuses
  • 1 operating instructions
  • 2 carbon brushes
  • 1 Notes on moving the equipment safely

Instructions for use

Allows to separate blood from serum/plasma and to get deposits of urinary samples or other puncture liquids samples.

Not suitable for tuberculosis cultures which require a "biosafety" centrifuge (with aerosol tight cover in case of tube breakage).

Always centrifuge the tubes with their cap in order to prevent any projection of liquid.

Always place the centrifuge on a level surface.

For round-bottomed tubes no rubber inserts are necessary.

Maintenance

Fuses must be replaced as required (ref: E891).

Carbon brush (ref: 0518) change must be carried out by authorised and qualified personnel.

MSF requirements

Basic model.

Essential in a reference laboratory.