AUTOCLAVE 24 l (All American), w/o burner, int. Ø 31.5 cm

STD ESTEAUTE24-

Valid Article

Former Code(s): ESTEAUTOA24
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
Z12011305 - Small load autoclaves with a chamber smaller than 1 standard sterilisation unit
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
The product is part of at least one Kit. A kit is a collection of products (medical and/or logistic) that are needed for a certain intervention in emergency. The choice and quantity of the articles reflects the MSF protocols for this specific situation. The use of Kits allows to start an intervention without a detailed evaluation.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

AUTOCLAVE (All American)

Definition

A non-electrical device designed for total elimination and/or inactivation of microorganisms from medical devices and related products using pressurized steam (i.e., moist heat) as the sterilizing agent.

Synonym

pressure cooker

Specifications

  • Aluminium vessel with handles
  • Aluminium lid with
    • handle
    • pressure gauge
    • regulating valve (or control /escape valve) extended by a flexible metal tube
    • overpressure plug and safety valve
  • Non perforated aluminium basket with feet
Image
Some restricted information has been hidden. Sign in to see this information

Quality standards

  • EN 13060, 2014, +A1 2018 Small steam sterilizers
  • EN 13445, 2021, Unfired Pressure Vessels​ (parts 1-5)
  • EN IEC 61010-2-040, 2020, edition 3, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-040 Particular requirements for sterilizers and washer-disinfectors used to treat medical materials
  • ISO 17665-1, 2006, edition 1, (confirmed 2016) Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices

Technical specifications

  • No lid seal: the joint is metal-to-metal.
  • The flexible metal tube enables pockets of air staying at the bottom of the vessel to escape.
  • Pressure gauge graduated in kg/cm2, bar, PSI, T ºF.
  • Although graduated up to 2 bars, this unit can only be used for sterilization at 1 bar = 121ºC = 15 PSI.
ESTEAUTE24-ESTEAUTE39-
Gross capacity24 litres39 litres
Overall height42.5 cm48.3 cm
Bottom height31.2 cm36.2 cm
Internal diameter32.1 cm38.7 cm
Aluminium container
  • Inside depth

26 cm

26.7 cm
  • Inside diameter

28.3 cm

35.6 cm

  • Capacity

16.6 liter

26.4 liter

Weight

  • Shipping weight

11.8 kg

19.5 kg
  • Unit weight

8.3 kg

15 kg

Packaging & Labelling

Unit packaging: reinforced cardboard box.

To be Ordered Separately

  • a heat source (kerosene stove, electric heating plate 1,500 W min… to be purchased locally if possible)
  • a timer
  • TST indicators
  • autoclave steam indicator tape
  • tripod
  • vaseline to lubricate the cover

Instructions for use

Some restricted information has been hidden. Sign in to see this information

Precautions for Use

  • When removing the cover, always tilt and angle the cover away from yourself or any other people in the area to prevent injury from hot steam.
  • Never heat the autoclave without water, before carrying out sterilization once more, the autoclave should be refilled with water but never reuse the water held in the removable lower tank since it is particularly limestone concentrated.
  • Never pour cold water into or onto the warm autoclave!
  • Use FILTERED water, non salt, low or non chlorinated, and preferably soft water such as rainwater
  • Place the autoclave on a stable surface: have a tripod made locally in wrought iron, with retaining lugs round the surface on which the autoclave sits.

Maintenance

  • Remove limescale build-up from time to time by heating the autoclave with vinegar diluted with water.
  • Never use chlorinated products to clean the autoclave.
  • Metal-to-metal seal
    • Must be lubricated periodically: apply a small amount of lubricant every third or fourth use.
    • Wipe off the seal with a clean towel to remove any build-up of foreign material or particles trapped in the lubricant.
  • Pressure gauge
    • Do not immerse the pressure gauge in water when cleaning.
    • Make certain the opening into the gauge on the underside of the cover is open and free of any foreign matter.
  • Control valve: periodic cleaning is recommended. Unscrew the “knurled top” portion and clean thoroughly in hot soapy water.
  • Air exhaust tube:
    • Must be frequently checked to make sure that air passes freely through it: blow air through the tube at least once a month.
    • Clean out the tube by using a small diameter wire, running it through the entire length of the tube several times.
    • If corrosion or a buildup of foreign material is noticed, the tube needs to be replaced.
  • Excess pressure valve: to be replaced every three years in normal service.
  • The overpressure plug (silicone, red in colour, on the top surface of the sterilization cover)
    • It is recommended to replace the overpressure plug every 6 months. It should always be replaced whenever it becomes hard or deformed.
    • At least every month, during period of use, the opening in the cover where the overpressure plug fits should be checked (no foreign material residue, buildup of grease) and cleaned with hot soapy water (additional to the daily cleaning).

Storage

If the autoclave is not going to be used again, before putting the unit away, all water should be emptied from the unit and the unit be thoroughly dried inside. It is recommended that the water be poured out while the bottom is still warm. For storage purposes, it is only necessary to slightly tighten the wing nuts enough to hold the cover on the vessel. When storing, it is recommended that the control valve be left in a vertical position.

MSF requirements

Suitable for health centres and small hospitals (up to 100 beds), or for large hospitals, either to supplement a 90 l autoclave, or for decentralized sterilization in hospital departments.

This type of autoclave is distributed by UNICEF and is found nearly everywhere, so spare parts can be obtained more easily locally.