OXYGEN FACE MASK, non-rebreathing + tubing, paediatric

STD SCTDMASOH1P

Valid Article

Single patient multiple use A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
Article is intended to be used with children (<12 years) ONLY.
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
R03010206 - Masks with reservoir
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
The product is part of at least one Kit. A kit is a collection of products (medical and/or logistic) that are needed for a certain intervention in emergency. The choice and quantity of the articles reflects the MSF protocols for this specific situation. The use of Kits allows to start an intervention without a detailed evaluation.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

OXYGEN FACE MASK, non-rebreathing + tubing

Definition

A flexible, form-shaped device designed to be placed over the nose and mouth to deliver air of high oxygen concentration to a patient's airway. It includes two unidirectional valves, one that closes during inspiration to prevent room air mixing with oxygen in a reservoir bag, and one that closes during exhalation to prevent exhaled respiratory gases from entering the reservoir bag.

It is a soft, flexible mask with an airtight seal against the patient's face. The device includes a headstrap and tubing.

Synonym

High concentration oxygen mask, mask with reservoir bag

Specifications

Quality standards

  • EN 13544-2, 2002, +A1 2009 Respiratory therapy equipment. Tubing and connectors
  • ISO 15001, 2010, edition 2, (confirmed 2022) Anaesthetic and respiratory equipment — Compatibility with oxygen

Technical specifications

  • Mask
    • flexible transparent PVC
    • exhalation valves with flexible membrane on both sides of the mask, enabling outflow of exhaled air (CO2) and preventing inflow of ambiant air
    • nose clip is soft, malleable and adjustable
    • elastic fastening strap (latex-free)
  • Reservoir bag
    • flexible and translucent
    • capacity: +/- 750 ml for adult mask, +/- 400 ml for pediatric mask.
  • Non-return valve between the mask and the reservoir, preventing exhaled air from entering the reservoir
  • Oxygen tube
    • PVC low resistance tubing, round shape section, designed for low-flow procedures (0-15l/min)
    • Tubing compatibility with standard oxygen connecting tubing, 3-5mm internal diameter and 7-8mm external diameter, and 15/22 mm diameter ventilation tubing.
    • length ± 200 cm
    • can be disconnected from the mask to allow a quick shift from the oxygen tubing from the mask to the resuscitator
  • Non sterile, for single patient use

Packaging & Labelling

Unit packaging

Instructions for use

Some restricted information has been hidden. Sign in to see this information
Some restricted information has been hidden. Sign in to see this information

Precautions for Use

  • The flow rate of the oxygen source must be twice higher than the patient's minute volume otherwise the reservoir doesn't fill up and the patient cannot breath.
  • The oxygen concentrator should have a flow rate of minimum 6-8 litres/min. If greater than 10l/min of flow is required, then two 10l/min oxygen concentrators can be Y-connected to the non-rebreather oxygen mask. Order additional tubing and Y-connector (see related articles).
  • Flow rate must be sufficient to keep the bag 1/3 to ½ inflated at all times.
  • Avoid kinking and twisting of the reservoir.
  • Fit the mask properly on the patient's face not to decrease oxygen concentration.
Some restricted information has been hidden. Sign in to see this information