SHARPS CONTAINER, 1 - 2 l, plastic, s.u.

STD SINSCONT2P-

Valid Article

Former Code(s): -X
Single use
A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
A0180 - Needles - accessories not included in other classes
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
The product is part of at least one Kit. A kit is a collection of products (medical and/or logistic) that are needed for a certain intervention in emergency. The choice and quantity of the articles reflects the MSF protocols for this specific situation. The use of Kits allows to start an intervention without a detailed evaluation.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

SHARPS CONTAINER, plastic, disposable

Definition

A single use plastic container used for the safe deposit and collection of biologically or chemically contaminated used sharps; this includes (long) needles, lancets, scalpel blades, cannulae. This container is puncture resistant, leak-proof, and closable. It is disposed of when filled to the recommended level, according to the facilities' instructions.

Specifications

Quality standards

ISO 23907-1, 2019, edition 1, Sharps injury protection - Requirements and test methods - Part 1: Single-use sharps container

Technical specifications

  • Rigid plastic: the solidity allows them to withstand perforations, leaks and impacts due to falls
  • Cylindrical model for the SINSCONT03P and SINSCONT2P-
  • Useful capacity:
    • 0.3 - 0.6 l (SINSCONT03P)
    • 1 to 2 l (SINSCONT2P-)
    • 4 l (SINSCONT4P-)
  • Anti-reflux system: protects against overturning in the event of accidental tilting of the collector and against unauthorized removal of soiled objects.
  • Sharps waste aperture with:
    • Temporary safety closure between eliminations or during transport
    • Tamper-proof closing
    • Uncoupling notches for Luer
  • Solid handle allowing easy transportation
  • Marking on lateral sides:
    • pictorial instructions for use
    • logo + manufacturer's name and address
    • international bio-hazard symbol
    • horizontal mark indicating the maximum fill line which is between 75 and 85% of the maximal volume

Packaging & Labelling

Unit packaging

Instructions for use

Context of use

For mass immunization campaigns, the 15 litres Safety Sharps box accommodating larger waste volumes is preferred.

In non-emergency situations, MSF recommends to use reusable sharps waste containers (see related articles below).

This container is not suitable for contaminated medical waste (compresses...).

Some restricted information has been hidden. Sign in to see this information

Waste management

SINSCONT03P and SINSCONT2P-

Once filled, the containers must be disposed of in the sharps pit, or incinerated in a specialized medical waste incineration plant with automatic loading and offloading systems

SINSCONT4P-

Once filled, the containers must be burnt with additional combustibles (like wood) in a Safety Box Reducer if not containing glass, or incinerated in a specialised medical waste incineration plant with automatic loading and offloading systems.

If these automatic systems are not available, special care and precautions must be taken. Contact your water & sanitation / environmental health technical referent at headquarter.