(EZ-IO) TRANSPORT BAG for EZ-IO kit, ref 9065
STD
SINSNEIOK11
Valid Article
Account code:
60210
HS Code:
392390
Last Updated on:
22/10/2024, 13:05:11
Former
Code(s):
-X
The product is part of at least one Kit.
A kit is a collection of products (medical and/or logistic) that are needed for a certain intervention in emergency. The choice and quantity of the articles reflects the MSF protocols for this specific situation. The use of Kits allows to start an intervention without a detailed evaluation.
INTRA-OSSEOUS INFUSION SYSTEM: EZ-IO DRILL
Definition
Battery powered medical drill and sterile medical devices needed to perform an intraosseous (IO) infusion (i.e., an infusion directly into the bone's medullary cavity) in children or adults.
Intraosseous infusion is used in emergency care when peripheral vascular acces is difficult, impossible or uneffective; especially for patients for whom immediate vascular access is lifesaving (e.g. patients with shock, severe dehydration, burns or cardiac / respiratory arrest)
Specifications
Some restricted information has been hidden. Sign in
to see this information
Technical specifications
BATTERY POWERED MEDICAL DRILL (ref. 9058)
- hand-held waterproof power driver
- powered by sealed lithium battery
- single use battery that cannot be replaced. The battery indicator (LED) will blink red when the trigger is activated and has only 10% of battery life remaining.
- life expectancy: 10 years or +/- 500 insertions
- no maintenance required
- dimensions : 16.5 cm x 11.4 cm x 6.4 cm
- weight: 315 g
TRANSPORT BAG for EZ-IO kit (ref 9065)
- soft sided
- yellow
Transport Dangerous Goods
- UN3091: Lithium metal batteries packed with equipment
- Class: 9
- Packing instructions: 970 Section II
- Limit per package:
- Pax A/C (Passengers & Cargo Aircraft) = 5 kg
- CAO (Cargo Aircraft Only) = 5 kg
Instructions for use
The drill comes with instructions for use, to be consulted before applying.
Precautions for Use
- Do not immerse the apparatus when performing cleaning and disinfecting.
Maintenance
- The exterior surface of the drill can be cleaned and disinfected with a detergent/disinfectant for surfaces/non-invasive medical equipment. Prepare use-solution in concentration required. Always read the label and product information before use. Do not mix with other products. Make sure to wet the surface completely, use a wipe or towel, keep it wet for the whole exposure time.
- Clean and disinfect also the trigger of the apparatus with the same detergent-disinfectant solution.
- Using swabs, moistened with the detergent-disinfectant solution, gently clean the inside opening around the metal drive shaft.
- After cleaning, inspect to ensure no visible debris remains, and no damage has occurred.
- Dry the driver with a soft, clean cloth and store protected from humidity and dust (use the transport bag).
(Cf Introduction: Disinfection and sterilization in the field)
Some restricted information has been hidden. Sign in
to see this information
Some restricted information has been hidden. Sign in
to see this information