(Erba XL200) XL MULTICAL 4x3ml, XSYS0034
STD
ELAECCHC726
Valid Article
HS Code:
901890
Last Updated on:
25/12/2024, 22:01:36
Former
Code(s):
SLASZNL0212
W0101050301 - Calibrators multicomponent (cc)
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
Thermosensitive codes are defined for storage and transportation temperature requirements of the products.
(Erba XL200) XL MULTICAL, XSYS0034
Definition
Lyophilized calibrator based on human serum used for calibration of clinical chemistry assays on Erba XL200 chemistry analyzer.
Specifications
Some restricted information has been hidden. Sign in
to see this information
Technical specifications
Light yellow solid
Packaging & Labelling
4 x 3 ml
To be Ordered Separately
- Quality controls
- Reagents
Instructions for use
Some restricted information has been hidden. Sign in
to see this information
Storage
- Unreconstituted lyophilized calibrator is stable till the expiry date mentioned on the label when stored at 2–8°C.
- Stability of the reconstituted calibrator*:
- at 15–25 °C 8 hours
- at 2–8 °C 2 days
- at (-25) – (-15) °C 4 weeks (when frozen once)
*Exceptions:
- Stability of Total Bilirubin in the reconstituted calibrator (when stored protected from light):
- at 15–25 °C 6 hours
- at 2–8 °C 1 day
- at (-25) – (-15) °C 2 weeks (when frozen once)
- Stability of Direct Bilirubin in the reconstituted calibrator (when stored protected from light):
- at 15–25 °C 3 hours
- at 2–8 °C 8 hours
- at (-25) – (-15) °C 2 weeks (when frozen once)
- Store calibrator tightly capped and protected from light when not in use. Store vials upright to avoid spilts and leakage. Seal the vials tightly after use to prevent evaporation.
MSF requirements
Use with the Erba XL200
Some restricted information has been hidden. Sign in
to see this information