FILM DRESSING, semi-permeable, adhesive, 10cmx10m, roll

STD SDREFIDSR1010

Valid Article

Former Code(s): SDREZBE0065 SDRETAPA1T0 SDRETAPA1S0 SDREFIDS10R10 SDREZNL0071
Single use
A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
M04010202 - Polyurethane fixing dressings
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
Thermosensitive codes are defined for storage and transportation temperature requirements of the products.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

FILM DRESSING, semi-permeable, adhesive, roll

Definition

A non-sterile, transparent, semi-permeable (i.e., impermeable to fluids, permeable to vapours and gases) covering applied to skin typically intended to fix a primary dressing or a medical device, or to provide protection from fluids. It is a thin, clear film made of synthetic polymer material that is covered on one side with a pressure-sensitive adhesive. This is a single-use device.

Specifications

Quality standards

EN 13726-2, 2002, Test methods for primary wound dressings. Moisture vapour transmission rate of permeable film dressings

Technical specifications

  • transparent film made of synthetic polymer (e.g., polyurethane based), latex- and PVC-free
  • breathable: allows moisture vapour and gas exchanges
  • moisture vapour transmission rate ≥ 500g/m²/24h
  • waterproof: impervious to liquids and body fluids
  • protection against all external contamination
  • barrier against bacteria and viruses ≥20nm
  • adhesive:
    • must adhere to the skin during 7 days minimum
    • hypoallergenic latex-free
  • without anti-microbial agents
  • easy to be cut into shape
  • non sterile, roll presentation
  • sizes
    • length: +/- 10m
    • width:
      • SDREFIDSR1510: 15 cm
      • ​SDREFIDSR1010: 10 cm

Packaging & Labelling

Instructions for use

Precautions for Use

  • To ensure adequate dressing adhesion, the skin around the wound / puncture site should be clean, dry and free of moisturisers and grease.
  • Caution is required when removing the dressing: stretch the film horizontally away from the wound to avoid damage to the skin.

Storage

  • Store below 25°C
  • Protect from sunlight ‐ Protect from humidity

MSF requirements

Transparent polyurethane film dressing, easy to cut, allowing the protection of a wound or puncture site against liquids and colonization by germs from the outside

Some restricted information has been hidden. Sign in to see this information