MICROBIAL DECTECTION SYSTEM (BacT/ALERT 3D 60)
Valid Article
BacT/ALERT 3D
Definition
The BacT/ALERT® 3D Microbial Detection System is an automated test system capable of incubating, agitating, and continuously monitoring aerobic and anaerobic media inoculated with patient specimens suspected of having bacteremia, fungemia, and/or mycobacteremia.
Specifications
Quality Standards Comment
The BacT/ALERT ® 3D Microbial Detection System conforms to :
- Relevant European regulations for electrical safety (UL-61010-1; IEC 61010-1; IEC 61010-2-081; CAN/CSA-C22.2 No. 1010.1-92);
- Electromagnetic compatibility (EMC) regulations.
Components
The BacT/ALERT® 3D system includes a single instrument containing an incubation chamber, monitor, keyboard, UPS, printer, barcode reader, and a mouse.
- BacT/Alert 3D 60: one incubation chamber containing 3 racks, each with a capacity of 20 culture blottles.
- Bact/Alert 3D 120: capacity of 120 culture bottles (2 drawers).
Technical specifications
Power supply:
- Rated AC voltage range: 100 - 240 V~, 50-60 Hz
- Electrical ground required
Operational environmental parameters:
- Operating temperature: 10°C to 30°C
- Relative humidity: 10 to 90%, non-condensing
- Altitude: 2000 meters
Storage:
- Temperature: -17°C to 57°C
- Relative humidity: 10 to 90%, non-condensing
Dimensions
BacT/ALERT 3D 60:
- Width — 57.8 cm
- Height — 61 cm
- Depth — 49.5 cm
- Weight (Unloaded) — 41 kg
- Weight (Loaded) — 45 kg
Instructions for use
The following specimen types can be used in the system:
- Blood
- Sterile body fluids
Precautions for Use
- All biological samples must be considered as potentially infectious and handled with the usual precautions (compulsory wearing of gloves, hand washing, etc.).
- Once placed in the unit, handling of specimen bottle is not required until a result is obtained.
- If microorganisms are present in the test sample, carbon dioxide is produced as the microorganisms metabolize the substrates in the culture medium. When growth of the microorganisms produces CO2, the color of the sensor in the bottom of each culture bottle changes from dark to light.
Maintenance
Routine BacT/ALERT ® 3D preventive maintenance is determined by your bioMérieux service program. There is no additional requirement for customer performed preventive maintenance. Contact your local bioMérieux Representative to schedule periodic service.
Waste management
After use, sterilize the cultures and all contaminated material.
MSF requirements
Must be validated by the laboratory referent