TEMP. LOGGER (Blulog TDL2-3Y) for world except EMEA

NST PCOLMONISBTS

Valid Article

OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

TEMP. LOGGER (Blulog TDL2-3Y) for world except EMEA

Form

  • Temperature scale: °C
  • Temperature range: -40 °C to +60°C
  • Average accuracy: 0.3 °C
  • Resolution: 0.1 °C
  • Measurement interval: 10 min
  • Memory capacity: 5,000 readings with automatic rollover
  • Radio frequency: 920.9 MHz
  • Protection class: IP67
  • Encryption: AES128
  • Time: UTC
  • Equipment needed: Hub
  • Power supply : internal battery
  • Period of validity: 3 years and 3 months
  • Data reading period post end of recording: 1 year
  • Range in building: ~30 -100 meters
  • Weight: 30 g
  • Dimensions: 85x55x5mm

Closed article. Per piece.

Fit

Each logger is provided with a holder and double-face adhesive for easy wall installation.

Image

Function

TDL2-3Y loggers are devices are used for measuring and sending temperature data every 10 minutes to a gateway (hub) through radio frequency.