KIT ComASP COLISTIN, clsi, eucast [Liofilchem-75001]
Valid Article
KIT ComASP COLISTIN
CAUTION
This test is worded and codified by unit to make the order easier, but it is always packaged in kit of 16.
Definition
System for colistin susceptibility testing with the broth microdilution method.
ComASP Colistin is a 4-test panel containing the dried up antibiotic in 7 two-fold dilutions (0.25 - 16 μg/ml).
The system is used to perform the broth microdilution (BMD) method for the antimicrobial susceptibility testing of colistin (polymyxin E) as recommended by international standards (i.e. CLSI, EUCAST) .
Specifications
Components
- 4 Systems (panels) of ComASP Colistin (panels individually packed in foil with silica gel desiccant)
- 16 Vials of Mueller Hinton II Broth (3.6 ml)
- Sealing Film
- Instructions Sheet and Test Results Form
Packaging & Labelling
4 x packs with 4 panels in the same pack
Instructions for use
- A panel allows the susceptibility testing of 4 different isolates.
- All wells in a single row (A, B, C or D) are rehydrated with a standardized microbial suspension.
- After incubation in thermostat the result is read and interpreted.
Precautions for Use
ComASP Colistin is a disposable device to be used only for diagnostic use in vitro. The product must be used in the laboratory by properly trained personnel, using approved aseptic and safety methods for handling pathogenic agents.
Storage
- Store ComASP Colistin at 2-8°C in the original packaging.
- Once an envelope is opened the panel should be used within 7 days.
- Keep away from direct sunlight and direct heat.
- Do not use the panels beyond the expiry date indicated on the label.
- Eliminate without using if there are signs of deterioration.
- Shelf life: 15 months
- Guaranteed minimum remaining shelf life at delivery: 1/3 of total shelf life
Waste management
After use, ComASP Colistin and material that has come into contact with the sample must be decontaminated and disposed of in accordance with guidelines used in the laboratory for decontamination and disposal of potentially infected material.
MSF requirements
Reserved for bacteriology culture programmes
This test can only be used with colistine in antibio which requires also approval of the referents for each order