HAND-HELD ULTRASOUND (Lumify) + probes S4-1, C5-2,
Valid Article
HAND-HELD ULTRASOUND (Lumify) + probes S4-1, C5-2
The technical specifications are in the process of being defined. For more information, contact the Diagnostic Imaging Working Group (diagnostic.network@amsterdam.msf.org).
Definition
The Philips Lumify is a portable handheld ultrasound device used to visualize internal body structures. Clinical uses include lung, cardiac, abdominal, pelvic, musculoskeletal, vascular (including venous access), nerve, soft tissue, and superficial imaging.
Specifications
Components
The following accessories are included in EDIMULSE5E-
- 1 Lumify C5-2 Curved Array Transducer (Abdomen, Gallbladder, Lung, OB/GYN)
- 1 Lumify S4-1 Broadband Sector Array Transducer (Abdomen, Cardiac, FAST, Lung, OB/GYN)
- 1 Samsung Galaxy S7 10.5'' tablet 5 year warranty
- 1 Samsung Galaxy S7 10.5'' tablet cover
Technical specifications
- Imaging features
- 2D
- Colour Doppler
- M-mode
- advanced XRES and multivariate harmonic imaging
- SonoCT
- The images and clips stored in the compatible smartphone or tablet can be exported via WiFi in DICOM format to a PACS, to a shared disk or by email in formatpng / mpeg.
- MicroB USB transducers with replaceable cable
- C5-2 Curved array transducer
- Bandwidth: 5-2 MHz
- Field of view: 67.3°
- Footprint: 50 mm
- Scan depth: up to 30 cm
- Applications: Abdomen, Gallbladder, Lung, OB/GYN
- Centre line marker
- S4-1 Broadband sector array transducer
- Bandwidth: 4-1 MHz
- Field of view 90°
- Footprint: 20.2 mm
- Scan depth: up to 30 cm
- Applications: Abdomen, Cardiac, FAST, Lung, OB/GYN
- Centre line marker
Instructions for use
Maintenance
The transducer is not autoclavable.
After each examination, the transducer must be cleaned and disinfected with a detergent-disinfectant solution for medical equipment (either 25 ml = 1 stroke of dosing pump for 5 liters of water).
(Cf Introduction: Disinfection and sterilization in the field)
Storage
- Temperature +10°C - +40°C
- Relative humidity: 15 - 95 %
- Atmospheric pressure: 700 - 1050 hPa
MSF requirements
This high-resolution ultrasound system offers similar performance features as larger cart-based systems in a portable, handheld, stand-alone unit.