OROPHARYNGEAL AIRWAY, s.u., sterile, 60mm ID 4.0mm black

STD SCTDAIRG060S

Valid Article

Former Code(s): EANEZPI0014

The product is supplied (and stored) sterile, it must remain sterile until its use. Sterile = state of being free from all living microorganisms.
Single use
A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
R010102 - Guedel airway tubes
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
The product is part of at least one Kit. A kit is a collection of products (medical and/or logistic) that are needed for a certain intervention in emergency. The choice and quantity of the articles reflects the MSF protocols for this specific situation. The use of Kits allows to start an intervention without a detailed evaluation.

OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

https://unicat.msf.org/web/image/product.template/587598/image_1920?unique=e82feed

OROPHARYNGEAL AIRWAY, disposable, sterile

Definition

A curved plastic tube inserted through the mouth to facilitate airway patency for gas exchange or suctioning. The device prevents the tongue from obstructing airflow. This rigid tube offers a bite-resistant passage for air from the lips to the posterior pharynx. This is a single-use device.

Synonym

oral airway, OPA, Guedel airway, Mayo cannula

Specifications

Quality standards

  • ISO 5364, 2016, edition 5, (confirmed 2022) Anaesthetic and respiratory equipment - Oropharyngeal airways
  • ISO 18190, 2016, edition 1, Anaesthetic and respiratory equipment - General requirements for airways and related equipment

Technical specifications

  • Design: Guedel
  • Material: plastic, latex-free
  • Parts:
    • curved and flattened distal (or pharyngeal) end
    • oval aperture, same lumen from proximal to distal end
    • Straight and reinforced proximal (or buccal) end​ = bite block
    • proximal flange (collar)
      • indicates the size by number and color code
      • is intended to be external to the teeth or gums
  • Sterile, for single use

​Different sizes: the size is identified on the flange of the cannula (ISO size = length in cm) and with the standard colour as defined by the ISO.

Packaging & Labelling

Quantity per box differs according the manufacturer.

Instructions for use

The proper size is selected by holding the airway against the side of the patient's face and ensuring that it extends from the corner of the mouth to the angle of the jaw. If the airway is improperly sized, it will occlude the airway.

An oral airway is placed by inserting a tongue depressor into the patient's mouth to displace the tongue downward and then passing the airway into the patient's mouth, slipping it over the patient's tongue.

In adults it is inserted into the mouth upside-down, and rotated into the normal position once it is inserted about half way.

Lubricant may be useful if the mouth is excessively dry.

Some restricted information has been hidden. Sign in to see this information

Precautions for Use

Do not resterilize

MSF requirements

The sterile cannulas have replaced the non-sterile ones as they are more available on the market.

Some restricted information has been hidden. Sign in to see this information

Extra Tables

Code

Use

Utilisation

Size ISO

Taille ISO

Length

Longueur

Internal diameter

Diamètre interne

Colour

Couleur

SCTDAIRG040S

Neonate

Nouveau-né

4

40 +/- 2.5 mm

3 mm

Pink

Rose

SCTDAIRG050S

Baby

Bébé

5

50 +/- 2.5 mm

3.5 mm

Blue

Bleu

SCTDAIRG060S

Small child

Petit enfant

6

60 +/- 2.5 mm

4.0 mm

Black

​Noir

SCTDAIRG070S

Child

Enfant

7

70 +5.0 / -2.5 mm

4.0 mm

White

Blance

SCTDAIRG080S

Adolescent

Adolescent

8

80 +/- 5.0 mm

4.5 mm

Green

Vert

SCTDAIRG090S

Adult

Adulte

9

90 +/- 5.0 mm

4.5 mm

Yellow

Jaune

SCTDAIRG100S

Large adult

Grand adulte

10

100 +/- 5.0 mm

5.0 mm

Red

Rouge

SCTDAIRGS120B

Large adult

Grand adulte

12120 +/- 5.0 mm5.5 mmpurple