(inf. pump Agilia VP) MAINTENANCE TOOLS Z179953+ USB cable
STD
EIMTINPE401
Valid Article
HS Code:
903300
Last Updated on:
06/12/2024, 22:01:16
(inf. pump Agilia VP) MAINTENANCE TOOLS
Definition
Kit to perform the preventive maintenance and the calibration of the infusion pump Agilia VP
Specifications
Some restricted information has been hidden. Sign in
to see this information
Technical specifications
VP Maintenance kit
- To calibrate and control the flow rate
- 60 g - 0.01 g balance - Z190957 (1)
- 50 ml beaker - Z190955 (2)
- To calibrate and control the door
- Agilia door calibration tool - Z067059 (3)
- To calibrate and control the pressure sensor
- Digital manomieter with RS232 - Z067060 (4)
- Serial-serial cable - Z190956 (5)
- USB to Serial converter cable - Z178031 (6)
- Screwdrivers
- Torx T10 screwdriver - Z700883 (7)
Agilia USB cable (Z073491)
To be Ordered Separately
- Agilia Partner Software - Z067049
- 3-way stopcock with Luerlock connector
- 20 ml syringe with Luerlock connector
Instructions for use
Some restricted information has been hidden. Sign in
to see this information
Precautions for Use
It is necessary to have training to use this kit.
Maintenance
As per Manufacturer recommendation the maintenance kit has to be checked by an accredited body
- every 6 months for (manometer + scale)
- every 12 months for the door tool .
MSF requirements
Maintenance kit compatible with the Agilia VP
Some restricted information has been hidden. Sign in
to see this information