BLOOD BANK REFRIGERATOR (B51), 220-240V, 24 bags

NST ELAEBBRE10-

Valid Article


CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
Z121701 - Blood banks refrigerators
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
The order of this product needs to be justified and is only acceptable under certain conditions.

OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

https://www.unicat.msf.org/web/image/product.template/587978/image_1920?unique=a7d5f92

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

BLOOD BANK REFRIGERATOR (B51), 230V, 45L (24 bags)

This blood bank refrigerator is currenlty undergoing field testing, before ordering contact your laboratory or biomedical advisor in order to put in place some monitoring.

Definition

Blood bank refrigerator is used to store blood products at a temperature of +4°C in a temperatures range between +2 and +6°C.

Specifications

Blood bank refrigerator with blood storage capacity of 24 units of blood bags of 450 ml.

Some restricted information has been hidden. Sign in to see this information

Quality standards

DIN 13277, 2022, Refrigerators and freezers for laboratory and medical applications - Terminology, requirements, testing

Material

  • Polypropylene
  • Aluminium

Technical specifications

  • Side door with glass
  • ​Volume: 52L, storage capacity: 45L
  • Blood bag capacity
    • 450 ml: 24
    • 350 ml: 31
  • Set temperature: +4°C
  • Alarm limits: +2°C & +6°C
  • Cooling technology: Forced air cooling system
  • Refrigerant type: Natural R600a 14g
  • Hold over time: 1.5h (+4°C to 10°C)
  • Digital display
  • Lock with 2 keys
  • Noise level: 35 dB
  • Natural defrosing technique
  • Power:
    • frequency: 50-60Hz
    • voltage: 220-240V
    • consumption: 0.8 kWh/24h
  • Environmental conditions
    • Maximum altitude: 3000m
    • Max humidity: 75%
    • Temperature range: 10-32°C

Dimensions

  • External dimensions (HxWxD): 670 x 495 x 575 mm
  • Internal dimensions (HxWxD): 470 x 375 x 350 mm
  • Net weight: 39 kg

Instructions for use

Read the users manual before use

Precautions for Use

  • Ensure that there is sufficient room around the unit for air circulation (min 70 mm)
  • Do not locate the unit near any heat source or direct sunlight.
  • Keep the unit in a dry well ventilated place.
  • Do not locate the unit below a ceiling fan or right next to air-conditioning equipment.
  • The unit must be transported in an upright position only (maximum inclination 45°).
  • To prevent problems with other electrical equipment causing this unit to malfunction, it should be connected to a separate circuit. Never connect the unit to a single socket with other electrical equipment by means of a multi-socket.
  • Secure all biomedical electrical devices in the laboratory. Contact your biomedical referent to define the best option.
  • Failing that, use a UPS double conversion for the electrical protection of the apparatus. Contact your biomedical referent for the best option.

Maintenance

  • Clean or disinfect the unit inside and outside before using it. Use only gentle cleaning agents. Never use aggressive or caustic cleaning agents, scouring powder, steel wool, abrasive sponges or chemical solvents (MSF Ref.: Surfanios).
  • When cleaning, make sure that no fluids of any kind run into the ventilation housing.
  • Read carefully the operator manual for more details on the cleaning and maintenance. All error codes are included in the manual.

Classification EC Regulation N° 1272/2008

  • H222 - Extremely flammable aerosol
  • H229 - Pressurised container: may burst if heated
Image

Signal Word

Danger

H222Extremely flammable aerosol.
H229Pressurised container: May burst if heated
P410 + P412Protect from sunlight. Do no expose to temperatures exceeding 50°C/ 122°F.
P210Keep away from heat/sparks/open flames/hot surfaces. – No smoking.
P211Do not spray on an open flame or other ignition source.
P251Pressurized container: Do not pierce or burn, even after use.
P101If medical advice is needed, have product container or label at hand.
P102Keep out of reach of children
P103Read label before use.

MSF requirements

Refrigerator complying to the minimal requirements for the storage of blood products, in the MSF field conditions. Supplied with a visible and audible alarm installed.

Some restricted information has been hidden. Sign in to see this information