THERMO-HYGROMETER tracer (LogTag Uhado-16) config. +15/+25°C

NST PCOLMONIHLU2

Valid Article

SC subscriptions: included in supply or field tools
OC subscriptions: included in supply or field order tools (e.g. UniField)
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)
https://unicat.msf.org/web/image/product.template/582091/image_1920?unique=97c3cb1

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

THERMO-HYGROMETER tracer (LogTag Uhado-16) config. +15/+25°C

Form

Humidity and temperature recorder that measures and stores up to 16000 sets of high resolution humidity and temperature readings over a measurement range of 0 to 100% RH and& -30°C to +70°C.

Recorder pre-configurated by the manufacturer (see at the end).

Specifications:

  • humidity resolution: better than 0.1% RH
  • temperature resolution: better than 0.1°C
  • capacity: 16000 pairs of humidity & temperature readings
  • memory: 32 KB
  • sampling frequency: adjustable, 30 seconds to several hour
  • data Logging: records time, humidity and temperature data for later download
  • day alarm summary: displays in calendar-like format alarm triggers for up to the last 30 days
  • current time, humidity and temperature: shows the current time and temperature when not in review mode
  • push button logging start with optional delayed start or specific time & date start
  • protection marking: IP61 (when hung or mounted vertically)
  • power source: 3V LiMg battery CR2032, no replaceable
  • battery life: 2 ~ 3 years of normal use*
  • dimensions: 93 mm (H) x 54.5mm (W) x 8.6mm (T)
  • weight: 41 g
  • case material: polycarbonate
  • optional clearing of alert indication by push button (an inspection mark is placed at same time)
  • every time the LogTag is downloaded the inspection mark recorded in its memory is downloaded as well
  • 'prestart' fail-safe logging (records temperature data even if not started)
  • low battery indication in software
  • comprehensive alert indicator configuration

MSF Logistique configuration for sea transport REFER container and refrigerated truck:

  • start with push button START
  • record readings continuously, overwrite oldest when memory full
  • recording frequency: 2 hours
  • starting delay after activation: 15 minutes
  • temperatures alarm: after 2 consecutive alert above +25°C or below +15°C
  • humidity alarm: after 2 consecutive alert above 65%
  • alarm remains on even if readings return to range +15/+25°C and beow 65%RH

Per piece.

Function

Recommended for temperature and relative humidity monitoring in sea transport REFER container and refrigerated truck.