VACUUM PUMP (ITE MK-280-DS) 2-stage, 230V 50Hz, 273l/min

NST CCLITOOLPI8-

Valid Article



OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

https://unicat.msf.org/web/image/product.template/582157/image_1920?unique=b6ddb0a

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

VACUUM PUMP (ITE MK-280-DS) 2-stage, 230V 50Hz, 273l/min

Form

A2-stage vacuum pump.

Specifications:

  • voltage: 230V, 50Hz
  • free air displacement: 273 l/min
  • ultimate vacuum: 12 micron
  • motor: 0.5kW
  • motor speed: 2880 rpm
  • oil capacity: 1070 ml
  • dimensions: 385 x 170 x 310 mm
  • net weight: 16,9 kg

This vacuum pump is equipped with the following:

  • electrically operated solenoid valve
  • vacuum gauge
  • oil mist filter
  • non return valve
  • dual voltage switch

Changed by - Ann-Sofie Vander Meulen now

Fit

This vacuum pump is suitable for use on all CFC, HCFC and HFC refrigerant systems.

Function

This 2 stage vacuum pump achieves deep vacuum and is used during installation or maintenance of ACs to remove air, moisture or contaminants from the refrigeration cycle to ensure efficient operation. Pulling vacuum is also used to detect leaks in the refrigeration system.