CLINICAL CHEMISTRY ANALYZER (Erba XL200), w/o ISE module
Valid Article
CHEMISTRY ANALYZER (Erba XL200), w/ISE module
Please contact your laboratory and biomedical advisor when considering the purchase of a clinical biochemistry automate, to ensure the appropriate choice, depending of your context.
Definition
Benchtop clinical chemistry automate with optional ISE module, measuring clinical chemistry parameters and ion-selective quantification of specific electrolytes (Na+, K+, Cl-, Li+) in serum, plasma, urine and cerebrospinal fluid (CSF).
Specifications
ISE: ion-selective electrodes
Automatic analyzer with additional ISE module intended for photometric quantification of clinical chemistry parameters and ion-selective quantification of specific electrolytes.
Components
- ISE module (optional)
 - Touch screen
 - Arm
 
Technical specifications
- For laboratories running 30-70 samples a day
 - Throughput:
 - 200 photmetric tests/hour
 - 400 ISE tests/hour
 - ISE simultaneous measurement: max. 45 photometric tests + 4 ISE
 - Sample: serum, plasma, urine, CSF
 - Analyzer compatible with vacuum tubes - centrifugation outside the device
 - Parameters: no limit on test parameters or calculated tests and 4 ISE parameters (Na, K, Cl, Li)
 - Reagent Tray: 50 refrigerated positions (8-12°C) 5, 20, 50 ml reagent containers
 - Sample Tray (39 positions):
 - Outer ring – 30 position for samples
 - Inner ring – 9 positions for blanks, standards, calibrators, controls and ISE solutions
 - Reagent Dispensing: One dispensing probe with liquid-level sensor with dispensed volumes: R1 50-300 μl – adjustable in 1 μl step; R2 10-200 μl – adjustable in 1 μl step
 - Sample volume: 2 - 70 µl
 - Reaction volume: 180 - 550 μl
 - Reaction Tray: 45 reusable hard glass cuvettes, optical path length 5 mm
 - PC requirements:
 - Operating system: Windows 10 Professionnal 32/64 bit, TBM OS of Windows 10 loT 32/64-bit,
 - Processor: Intel Core i3 (or higher)
 - RAM 4 GB (minimum)
 - HDD minimum 500 GB
 - Power supply: 220 V ± 10 %, 50 Hz ± 5%, 600 VA
 - Mixing system: Independant stirrer
 - Barcode reader: Built in barcode reader for samples and reagents
 - Water consumption: Max 7.5 l/hr
 - Working temperature: 15 to 30°C
 
Dimensions
- Dimensions: 810 mm x 700 mm x 600 mm
 - Weight: 120 kg
 
To be Ordered Separately
Please contact your laboratory referent for more informations concerning tests, controls, consumables, spare-parts or accessories.
Instructions for use
Precautions for Use
Secure all electrical devices in the laboratory with a UPS double conversion. To define the best sizing of the UPS, contact your biomedical or energy referent.
Storage
- Storage temperature: 5 to 45°C
 - Storage humidity: < 80% non-condensing
 
MSF requirements
Please contact your laboratory and biomedical advisor when considering the purchase of a clinical biochemistry automate.