(mb GeneXpert)TEST HIV-1 Qual XC EID,cart. GXHIV-QA-XC-CE-10
Valid Article
(mb GeneXpert)TEST HIV-1 Qual XC EID, cartridge
This article replaces and is the new version of ELAEMBIT104, with extented coverage.
CAUTION
This test (= 1 cartridge) is worded and codified by unit to make the order easier, but it is always packaged in kit of 10.
Definition
Xpert HIV-1 Qual XC (Extended Coverage) is an in vitro nucleic acid amplification test for the qualitative detection of human immunodeficiency virus type 1 (HIV-1) total nucleic acids, on the automated GeneXpert System. The test is used to detect HIV-1 in human dried blood spots (DBS) and EDTA capillary or venous whole blood (WB) specimens from individuals suspected of HIV-1 infection.
Xpert HIV-1 Qual XC is intended to aid in the diagnosis of HIV-1 infection in conjunction with clinical presentation and other laboratory markers in infant, adolescent and adult populations.
The test is not intended to be used as a blood donor screening test for HIV-1.
Specifications
Technical specifications
The Xpert® HIV-1 Qual XC (GXHIV-QA-XC-CE-10) contains sufficient reagents for 10 specimens. Each kit contains following:
- 10 HIV Qual (GXHIV-QA-XE-CE-10) assay cartridges with integrated reaction tubes
- Bead 1, Bead 2, and Bead 3 (freeze-dried) 1 of each per cartridge
- Lysis Reagent (Guanidinium Hydrochloride) 1.2 mL per cartridge
- Rinse Reagent 0.5 mL per cartridge
- Elution Reagent 1.5 mL per cartridge
- Wash Reagent (Guanidinium Hydrochloride) 3.2 mL per cartridge
- Proteinase K Reagent 0.48 mL per cartridge
- Disposable 100 μL Transfer Pipettes 1 bag of 10 per kit
- CD
- Assay Definition File (ADF)
- Instructions to import ADF into software
- Instructions for Use (Package Insert)
Beads are solid white components in cartridges; reagents are clear liquids which are primarily buffered in aqueous solutions
Quality control
Each test includes a Sample Volume Adequacy (SVA) control, a Sample Processing Control (SPC) and a Probe Check Control (PCC).
External controls, not available in the kit, should be used in accordance with Ministry of Health requirements as applicable.
Packaging & Labelling
1 kit = 10 tests
To be Ordered Separately
- GeneXpert DX System
- Bleach or hypochlorite sodium
- Ethanol or denaturated ethanol
If use of DBS:
- The Eppendorf ThermoMixer C and Smartblock need to be ordered for DBS application only (see related articles).
- DBS filter paper cards
- Blood sample collection items
- Sterile tweezers or forceps
- Sterile scissors
Instructions for use
Precautions for Use
- Do not open a cartridge until you are ready to perform testing
- Do not use the cartridge if it appears wet or if the lid seal appears to have been broken.
- Do not use a cartridge that has been dropped after removing it from the packaging.
- Do not shake the cartridge. Shaking or dropping the cartridge after opening the cartridge lid may yield invalid results.
- Do not use any reagents that have become cloudy or discolored.
- Do not use a cartridge that has leaked.
Storage
- Store between 2 and 28°C
- Shelf life: 18 months
- Guaranteed minimum remaining shelf life at delivery: 1/3 of total shelf life
Waste management
The cartridge contains guanidium thiocyanate as the lysis reagent. This chemical is highly toxic and all care should be taken when using the cartridges in case of spills etc.
Please note that as this chemical is also highly toxic to aquatic life, it should optimally be disposed of by incineration. No liquids from the cartridge should be released into the environment.
Classification EC Regulation N° 1272/2008
- Skin irritation, category 3
- Eye irritation, category 2
- Respiratory sensitization, category 1
Signal Word
Danger
H315 | Causes skin irritation |
H334 | May cause allergy or asthma symptoms or breathing difficulties if inhaled. |
H320 | Causes eye irritation |
P261 | Avoid breathing dust/fume/gas/mist/vapours/spray. |
P264 | Wash … thoroughly after handling |
P280 | Wear protective gloves/protective clothing/eye protection/face protection |
P321 | Specific treatment (see ... on this label) |
P362 | Take off contaminated clothing and wash before reuse |
P302 + P352 | IF ON SKIN: wash with plenty of soap and water |
P304 + P340 | IF INHALED: Remove to fresh air and keep at rest in a position comfortable for breathing. |
P305 + P351 + P338 | IF IN EYES: rinse cautiously with water for several minutes. If the victim wears contact lenses, remove them if possible. Keep rinsing |
P332 + P313 | If skin irritation occurs: get medical advice/attention |
P337 + P313 | If eye irritation persists: get medical advice/attention |
P342 + P311 | If experiencing respiratory symptoms: Call a POISON CENTER or |