PUMP, SUCTION, MECHANICAL (Twin Pump) + collection bottles

NST EHOEPUME1--

Outdated Article

Former Code(s): EHOEPUMS1B- EEMDPUME1-- EHOE1--
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Add article to list

   

Add Kit To Wishlist

   

Documents

PUMP, SUCTION, MECHANICAL (Twin Pump)

The twinpump is no longer produced by Ambu and replaced by the Fazzini pump F-170.

Definition

A portable assembly of devices primarily intended to be used by emergency medical services (EMS) to aspirate fluids, secretions, or other foreign materials by means of suction.

It typically consists of a manually-powered (hand or foot-operated) mechanism to drive the suction pump, tubing and a collection container. The pump creates a vacuum in the suction tubing, which is used for the removal of materials into the collection container.

Image

Specifications

The TWIN PUMP is composed of 1 suction pump and 1 suction tube, to which MSF added:

  • 1 x 10 m of silicone tube (SCTDTUBE081)
  • 2 collection bottles (EHOEPUMA110)
  • 10 biconical connectors (SCTDCONN7S)
Some restricted information has been hidden. Sign in to see this information

Quality standards

ISO 10079-2: 2014: Medical suction equipment -- Part 2: Manually powered suction equipment

Technical specifications

  • SUCTION PUMP
    • fitted with a double-piston pump allowing continuous aspiration
    • can be operated by hand or by foot (pedal)
    • suction capacity
      • max. vacuum: - 800 mbar (- 80 kPa /- 600 mm Hg)
      • free air flow at two pump actions per second: 30 - 40 litres/min
      • free air flow at maximum working rate of pump: ± 70 litres/min
    • dimensions: 206 x 96 x 104 mm
    • collection jar capacity: ± 600 ml, that is small and that is why a bottle was added
    • all parts are autoclavable up to 134º C
  • SUCTION TUBES
    • the first tube which connects the pump to the bottle is part of the pump
      • silicone
      • internal diameter: 10 mm
      • fitted with a fine nozzle
      • autoclavable
    • the second tube (SCTDTUBE08-) was added for connecting the bottle to the patient
      • silicone
      • internal diameter: 8 mm
      • supplied in 10 m roll to be cut according to the required dimensions
      • autoclavable
  • COLLECTION BOTTLE
    • glass
    • capacity: 2 litres
    • graduated
    • fitted with a latex plug with 2 antenna's + VIRPLAST ring to fix the plug + standard connector
    • autoclavable

Instructions for use

This item is supplied with very complete instructions for use.

Provide some spare collection bottles.

Yankauer cannulas have to be ordered separately (see related articles below).

Maintenance

The collection bottle must be pre-disinfected, cleaned and sterilized by steam autoclave after each use

The whole device can be cleaned and sterilized by steam autoclave.

(Cf Introduction: Disinfection and sterilization in the field)

After dismantling and cleaning, the pump should be tested to make sure that it works correctly.

MSF requirements

Emergency mechanical aspirator with a high aspiration capacity, which can be operated either by hand or by foot.

​Device allowing to perform aspirations in surgery/ anaesthesia when there is no reliable source of electricity.

Adding the collection bottle allows to increase the collection capacity of the pump.

Some restricted information has been hidden. Sign in to see this information