SYRINGE PUMP (Perfusor® compact), single syringe
Outdated Article
SYRINGE PUMP (Perfusor® compact)
Stopped in 2018 and becomes NST because no longer manufactured. Replacement by SYRINGE PUMP (AGILIA SP MC Z018610), single syr., drug libr. EEMDSYPE41- (see “replaced by” at bottom of the page).
Definition
Electrical device allowing infusing continuously and at constant flow rate a solution or a medicine contained in a syringe.
Synonym
electrical syringe
Specifications
Quality standards
- EN 60601-2-24, 2015, Medical electrical equipment - Part 2-24: Particular requirements for the basic safety and essential performance of infusion pumps and controllers
- ISO 8536-8, 2015, edition 2, (confirmed 2021) Infusion equipment for medical use - Part 8: Infusion sets for single use with pressure infusion apparatus
Components
EEMDSYPE1B- Perfusor® compact Plus: with the drug library DoseGuard™ included
Technical specifications
- Hard plastic case
- Metal handle
- Push arm with release lever and lock to maintain the syringe in place
- LCD display (backlit if plugged into the mains) displaying the following parameters:
- infusion rate
- type of syringe
- power supply
- witness of infusion
- alarms
- Programming keypad
- Programmable functions:
- total volume to be infused
- modification of the flow rate
- bolus function with simultaneous indication of the delivered volume
- pause
- For 20 ml and 50/60 ml Luer-Lock syringe
- Visual and sound alarms
- Power supply:
- mains: 230–240 V AC, 50/60 Hz
- with 4 batteries, AA/LR6 type
- thanks to the built-in rechargeable 12 V DC battery (autonomy: 20 hours)
- via a 12 V DC battery (with specific power supply cable)
Dimensions
190 x 100 x 120 mm
Transport Dangerous Goods
- UN3481 Lithium ion batteries contained in equipment
- Class: 9
- Packing Instructions: 967 Section II
- The label must include the UN code of the battery + an emergency number to contact in case of an incident
- Limit per package:
- Pax A/C (Passengers & Cargo Aircraft) = 5 kg
- CAO (Cargo Aircraft Only) = 5 kg
Instructions for use
To be used by duly trained staff only.
Always check that de calculation of the dilution and the registered rate correspond to the medical prescription.
Read carefully the manufacturer's leaflet and follow the instructions for maintenance.
Caution, not all brands of syringes are compatible with the syringe pump. When in doubt, please contact your supply centre.
Precautions for Use
Secure the electrical protection of the apparatus with a double conversion UPS (see related articles below).
Maintenance
Must be cleaned and disinfected with Surfanios® 0.25 % detergent-disinfectant solution (i.e. 20 ml = 1 sachet = 1 stroke of dosing pump for 8 litres water).
Do not rinse (contact time must be 15 minutes minimum and the product is not corrosive).