FIT TEST KIT, qualitative testing of FFP2/N95 respirators
STD
ELINMAFT1--
Valid Article
Account code:
60210
HS Code:
901890
Last Updated on:
22/11/2024, 22:19:15
Former
Code(s):
-X
RESPIRATOR FIT TEST
Definition
Kit allowing to qualitatively assess the adequacy of a FFP2 or N95 respirator by exposing the wearer to a test agent. If the wearer does not detect the presence of the test agent, the test is successful and respirator fit has been demonstrated.
This kit is used for:
- select the respirator which is best fitted to each user's face by carrying out fit testing with the supplied samples
- make sure that respirators are correctly worn by the concerned staff members by running a fit testing program
Specifications
Components of the set:
- One Hood (with collar assembly)
- Nebulizer nº 1 (Sensitivity Test)
- Nebulizer nº 2 (Fit Test)
- (Two Sets Replacement Nebulizer Inserts)
- Sensitivity Solution(s)
- Fit Test Solution(s)
Some restricted information has been hidden. Sign in
to see this information
Quality Standards Comment
Meets the performance criteria for fit testing respirators under the current OSHA Standard for Respiratory Protection: 29 CFR 1910.134.
Appendix A to §1910.134—Fit Testing Procedures (Mandatory)
Part I. OSHA-Accepted Fit Test Protocols
A. Fit Testing Procedures—General Requirements
Packaging & Labelling
Cardboard box with IFU
Instructions for use
Some restricted information has been hidden. Sign in
to see this information
Precautions for Use
- Make sure to order some boxes of respirators for the testing, different shapes and sizes.
- Be careful to avoid hyperventilation.
- If there is evidence of solution contamination, immediately discard the bottle.
Maintenance
- Rinse the nebulizers with warm water to prevent clogging and shake dry
- Wipe out the inside of the hood with a damp cloth or paper towel to remove any deposited solution
Storage
In a dry and well-ventilated place
Some restricted information has been hidden. Sign in
to see this information