FREEZER (Vestfrost MF314 G1) 323l, 230V

NST PCOLFREEVF3

Outdated Article

Former Code(s): PCOLFREE3E-
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

FREEZER (Vestfrost MF314 G1) 323l, 230V

Definition

Compressor chest freezer, used for storing temperature-sensitive products and freezing ice packs.

Specifications

  • power supply: 230 V AC
  • power: 250 W
  • electricity supply required at 43 ºC: minimum 8 hrs in every 24
  • gross storage capacity: 323 litre
  • ice pack freezing capacity: 32.4 kg in 24 hours
  • 0.6 litre ice pack storage capacity: 323
  • 0.4 litre ice pack storage capacity: 450
  • holdover time without power at 43 ºC: 17.5 hours
  • compressor type: Danfoss FR10.G OC
  • automatic thermostat
  • lockable lid
  • dimensions L x W x H: 156 x 60 x 85 cm
  • weight: 87 kg
Some restricted information has been hidden. Sign in to see this information

Instructions for use

It is vital to read the installation manual and follow the instructions precisely before starting the unit.

For all information concerning storage and handling of the products to be stored, refer to the "preliminary remarks" at the start of this family.

Affix a temperature recording form (ALSTTEMP41E) to each freezer.

Caution

A compressor freezer should be left to rest for 24 hours after transportation before being switched on, in order to allow the lubricating oil to drain back into the compressor.

Never tilt a compressor freezer by more than 45°.