FREEZER (Vestfrost MF314) 323l, 230V, upgrade

STD PCOLFREEV3U

Valid Article

Hazardous materials and products (identified by their 4-digit transport UN-number) must be transported under certain conditions as specified in the MSDS sheet (Material Safety Data Sheet).
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

FREEZER (Vestfrost MF314) 323l, 230V, upgrade

Image

Definition

Compressor chest freezer, used for storing temperature-sensitive products and freezing ice packs.

Specifications

  • power supply: 220 - 240 V AC
  • power: 179 W (4,29 kWH/24 hrs)
  • electricity supply required at 43 ºC: minimum 20 hrs in every 24 hrs
  • gross storage capacity: 281 litres
  • ice pack freezing capacity: 7.2 kg in 24 hrs
  • 0.6 litre ice pack storage capacity: 256
  • holdover time without power at 43 ºC: 4 hrs
  • automatic thermostat
  • integrated voltage stabilizer
  • refrigerant type: R600a
  • lockable lid
  • dimensions L x W x H: 73.5 x 163.5 x 95.5 cm
  • weight: 80 kg (106kg including packaging)
  • number of baskets: 3
  • 10 cm adjustable legs
Some restricted information has been hidden. Sign in to see this information

Transport Dangerous Goods

Classe 2.1 (flammable gas).

UN-code: UN3358 (REFRIGERATING MACHINES with flammable liquefied gas, non-toxic).

  • IATA Special Provision SP A103: less than 100gr gas can be transported by air
  • IMDG Special Provision DS291: can be transported by sea

Instructions for use

It is vital to read the installation manual and follow the instructions precisely before starting the unit.

For all information concerning storage and handling of the products to be stored, refer to the "preliminary remarks" at the start of this family.

Affix a temperature recording form (ALSTTEMP41E) to each freezer.

Caution:

A compressor freezer should be left to rest for 24 hours after transportation before being switched on, in order to allow the lubricating oil to drain back into the compressor.

Never tilt a compressor freezer by more than 45°.

Precautions for Use

Installation:

See manual for installation. In the user manual, it says that you have to install the appliance in a room with a volume of minimal 1m3 per 8g of hydrocarbon refrigerant used in the appliance, which is indicated on the rating plate. The required room volume should be:

Model

Quantities of R600a refrigerant

on rating plate [kg]

Minimal volume of room required (per freezer) [m3]
MF1140.058 kg7.25
MF2140.070 kg8.25
MF3140.075 kg9.40

Door seal:

Similar as the old models from Vestfrost, Vestfrost recommends to change the complete lid if the seal is old or broken. The seal alone is no longer avaiable.

Spare parts:

Spare parts can be ordered, some are compatible with the former models.

Gas:

R600a is a refrigerant gas which is a flammable. In case of leakage, the risks of explosion or fire are almost non-existent if the precautions for use have been respected (see user instructions for more details). In case of intervention on the refrigeration unit, make sure that the technician takes the necessary precautions and that an extinguisher is at hand.