REFRIGERATOR (Vestfrost MK144) 48l net, 230V

STD PCOLREFEVK1

Outdated Article

Former Code(s): PCOLFRID1E- PCOLREFEVK4
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

REFRIGERATOR (Vestfrost MK144) 94l

Definition

Note: this equipment has been outdated and can't be ordered. It was replaced by the Vestfrost refrigerator VLS200A. Spare parts for equipment in the field are still available.

Ice-lined compressor refrigerator used for storing temperature-sensitive products.

Specifications

  • Power supply: 230 V AC
  • Power: 160 W
  • Electricity supply required at 43 ºC: minimum 8 hrs in every 24
  • Net storage capacity: 48 litres
  • Gross storage capacity: 63 litres (vaccination campaign)
  • Number of baskets: 3
  • Number of ‘Water ballast, blue' needed: 2 (included with fridge)
    • Holdover time without power at 43 ºC: 37 hrs
    • Compressor type: Danfoss TLG5
    • Automatic thermostat
    • Lockable lid
    • Dimensions L x W x H: 60 x 92 x 85 cm
    • Weight: 72 kg (net) / 92 kg (gross)
    Some restricted information has been hidden. Sign in to see this information

    Instructions for use

    It is vital to read the installation manual and follow the instructions precisely before starting the unit.

    For all information concerning storage and handling of the products to be stored, refer to the "preliminary remarks" at the start of this family.

    Affix a temperature recording form (ALSTTEMP41E) to each refrigerator.

    Store products which should not be be frozen in a basket at the top of the refrigerator (at the bottom, temperatures are liable to fall below 0 ºC).

    Caution

    A compressor refrigerator should be left to rest for 24 hours after transportation before being switched on, in order to allow the lubricating oil to drain back into the compressor.

    Never tilt a compressor refrigerator by more than 45°.

    Some restricted information has been hidden. Sign in to see this information

    MSF requirements

    WHO/UNICEF standard