TUBE, SUCTION, conical tip, 50 cm, single use, CH16
STD
SCTDTUSU16-
Valid Article
Account code:
60210
HS Code:
901890
Last Updated on:
22/11/2024, 22:42:18
Former
Code(s):
-X
Single use
A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
R05010101 - Respiratory suction probes without a control hole
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
The product is part of at least one Kit.
A kit is a collection of products (medical and/or logistic) that are needed for a certain intervention in emergency. The choice and quantity of the articles reflects the MSF protocols for this specific situation. The use of Kits allows to start an intervention without a detailed evaluation.
SUCTION TUBE with conical tip
Definition
A flexible tube typically made of plastic designed for periodic aspiration of liquids and/or semi-solids from a patient's upper airway.
Specifications
Quality standards
ISO 8836, 2019, edition 5, Suction catheters for use in the respiratory tract
Technical specifications
- Transparent or translucent flexible PVC
- Straight, atraumatic, open distal end, with or without lateral eyes
- Proximal end with funnel shape connector
- Length: ± 40 to 60 cm
- Diameter: indicated by an international colour code:
size | ext diam mm | colour |
CH6 | 2.0 mm | light green |
CH8 | 2.7 mm | blue |
CH10 | 3.3 mm | black |
CH14 | 4.7 mm | green |
CH16 | 5.3 mm | orange |
- Sterile, for single use
Packaging & Labelling
Unit sterile packaging in peel-open pack
Instructions for use
Tube which is inserted through either mouth/nose or lumen of an endotracheal/tracheotomy tube, to clear food and secretions from upper airways or trachea.
This tube can be connected:
- either to a syringe with conical tip for manual aspiration
- or to a mechanical or electrical suction system (see related articles below)
The biconical connector for connecting the tube to the suction system should be ordered separately (see related articles below).