CLINICAL CHEMISTRY ANALYSER (Spotchem EZ)

STD ELAECCHE6--

Valid Article

Former Code(s): ELAEZNL0585
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
W0201010101 - Chemistry analyser - low routine (random access without ISE mode test/h <= 400)
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
The order of this product needs to be justified and is only acceptable under certain conditions.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

ANALYSER, CLINICAL CHEMISTRY (Spotchem EZ)

Definition

Dry chemistry analyzer for the quantitative measurement of biochemistry parameters from whole blood, serum or plasma samples using test strips.

Specifications

Some restricted information has been hidden. Sign in to see this information

Technical specifications

  • Specimen: serum, plasma or whole blood
  • Throughput: 63 tests/hour
  • Individual parameters or in panels (max. 9)
  • Warm-up time: 10 minutes
  • Measurement conditions: 10-30°C and 20-80% humidity
  • Processing Speed: 63 tests/hr
  • Minimum Sample: 6 x number of tests + 38µl (min. 250 µl when using whole blood in internal centrifuge)
  • Sample Container:
    • Whole blood (exclusive blood centrifuge tube)
    • Serum-plasma (exclusive serum sample tube)
  • Data Storage: 100 measurements
  • Calibration by magnetic card, Printer: 58mm width built-in thermal line printer (max. 36 characters / line)
  • Power Supply: AC 100-120V, 220 - 240V, 50/60 Hz (AC adapter method)

Dimensions

  • Weight: 5.4 kg
  • Dimensions: 338 mm x 203 mm x 167 mm

Supplied with the Article

  • Power cord
  • AC adapter
  • Accessory case:
    • 10x tips
    • Cleaning wire
    • Cleaner set: brush, cotton swab
    • Nozzle set: nozzle with O-ring
    • 2x wrench set for nozzle replacement
    • 10x centrifuge tubes
    • 2x tip waste cases
    • 2 protective covers
    • 1 thermal printer paper, width 58mm
  • Operating manual
  • Warranty card with case, delivery label

To be Ordered Separately

see related items:

  • Spotchem II reagents strips
  • for testing whole blood: whole blood sample tube (orange cap)
  • for testing serum plasma: serum sample tube (blue cap)

Instructions for use

Storage

Keep the device at 1-30°C and 20-80% humidity

MSF requirements

Please contact your laboratory advisor to help identify the best chemistry analyzer to answer your projects needs.

Some restricted information has been hidden. Sign in to see this information