(pulse oximeter Nellcor ) SENSOR EXTENSION CABLE
Valid Article
Nellcor SpO2 Cables
Definition
A non-patient-contact, insulated, electrical cord with a connector (plug) at both ends designed to transmit electrical power and/or signal (data) between medical devices (e.g., to connect a lead, electrical catheter, handpiece, or monitor to a parent device). It is not intended to connect directly to patient-worn electrodes (i.e., it is not a lead), is not intended to connect to the mains (i.e., not a mains power cable), does not generate any type of power and/or signal, and has no additional non-electrical conducting or processing functionality. Some types may include an electrical isolator to prevent electrical shocks. This is a reusable device.
Specifications
reusable pulse oxymetry cable, intended for use with all patient populations for continuous non-invasive arterial oxygen saturation (SpO2) and pulse rate monitoring when used with compatible host device and SpO2 sensor.
Product compatibility 2021406-001 and 2021406-002
- B850 with Nellcor OxiMax™ option with Tram module 451N5/851N5
- B650/B450/V100/VC150 with Nellcor OxiMax option
- Cardiolab/Mac-Lab with Tram module 451N5
- Dash 2500/3000/4000/5000 with Nellcor OxiMax option
- Solar 8000/8000M/8000i/9500 with Tram module 451N5/851N5
- S/5, AS/3, CS/3 modular with E-NSATX
Quality Standards Comment
In conformity with the requirements of the Medical Device Regulation 2017/745 and with the requirements of the RoHS Directive 2011/65/EU, article 4 on the restriction of the use of certain hazardous substances in electrical and electronic equipment.
Technical specifications
Interconnect cable with GE connector and Nellcor OxiMax Connector
Length cable 2021406-001 = 3m / 10 ft
Operating environment (= storage conditions)
- Temperature: -30° to 60°C
- Humidity: 10% to 90% RH
- Pressure: 70 kPa to 106 kPa
Materials Cables: Polyurethane (PU), Thermoplastic Polyurethane (TPU), and Polycarbonate (PC) (not made with BPA, natural rubber latex, PVC, DEHP or other phthalates.)
Packaging & Labelling
Packaging materials Primary: Polyethylene (PE)
Instructions for use
Maintenance
Cleaning solutions
When needed, spray with or wipe with pad moisturized with 90% ethyl alcohol or 60% isopropyl alcohol or 2% glutaraldehyde. Although discoloration may occur, <10% sodium hypochlorite, 0.75% tartaric acid, <5% quarternary ammonium compounds, or <2% phenol may be used.