MONITOR, multiparameter (Procare B40)+ accessories, 230V
Outdated Article
MONITOR multiparameter (PROCARE B40 V2)
The monitor Procare B40 V2 is no longer produced by GE Healthcare. The EMACC and biomedical groups are currently looking for a replacement.
In the meanwhile the B125 is the replacement monitor (see “replaced by” at bottom of the page).
Definition
Portable, vital signs monitor, used to provide continuous measurement of electrocardiogram, pulse oximetry and capnography, and discontinuous measurement of non-invasive blood pressure (NIBP).
Specifications
Quality standards
Components
- 1 monitor with colour screen = 26.4 cm display
- 1 power cable
- 2 Lithium-Ion batteries
- 1 instructions manual
Technical specifications
- Display:
- ECG graph (3 leads) + pulse rate
- expired CO2 waveform in mmHg
- values for SpO2 (%) and plethysmographic waveform
- values for non-invasive blood pressure (systolic, diastolic, and mean pressures)
- trend data
- Adjustable alarms: peripheral pulse rate, saturation, blood pressure, CO2, pulse and respiratory rates
- Power supply: 200-240 V/50 Hz (comes also in 110V/60 Hz) and Lithium-Ion battery (4 h ; 39 Wh)
Transport Dangerous Goods
- UN3481 Lithium ion batteries contained in equipment
- Class: 9
- Packing Instructions: 967 Section II
- The label must include the UN code of the battery + an emergency number to contact in case of an incident
- Limit per package:
- Pax A/C (Passengers & Cargo Aircraft) = 5 kg
- CAO (Cargo Aircraft Only) = 5 kg
Supplied with the Article
- ECG
- 1 multi-link ECG cable
- 2 clip 3-Ld ECG lead wires
- 3 boxes of 300 electrodes
- CAPNO
- 1 CO2 sensor, "sidestream" E-miniC
- 1 CO2 airway adapter adult/paediatric
- SpO2
- 1 SpO2 interface cable
- 2 reusable adult finger sensors
- 1 reusable neonatal sensor
- 2 x 100 adhesive tapes for neonatal sensors
- NIBP
- 2 NIBP dual hoses
- 9 BP cuffs (2 paediatric, 2 small adult, 3 adult, 2 large adult)
- Temperature
- 1 temperature cable 0.5 m (EEMDMONA103)
- 1 skin temperature probe Ø10mm, reusable (EEMDMONA125)
- 1 temperature probe general use, Ø3mm, reusable (EEMDMONA122)
Instructions for use
Device intended for the monitoring of patients in peri-operative activity or intensive care.
This capnography system must always be used with a hydrophobic antibacterial filter.
Operating conditions
- 0-40ºC, CO2 sensor 10-40ºC
- Relative humidity 5-95% at 4ºC
Precautions for Use
Secure the electrical protection of the apparatus with a double conversion UPS (see related articles below).
MSF requirements
The pulse oximeter and the capnograph are part of the basic anaesthetic equipment of any operating theatre.