CENTRIFUGE, electrical (Hettich 420), 230V 50-60HZ

STD ELAECENE17-

Valid Article

Former Code(s): Z ELAECENE17Z
W02069003 - Centrifuges
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
The order of this product needs to be justified and is only acceptable under certain conditions.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

CENTRIFUGE (Hettich 420)

Reserved for TB culture programmes.

Definition

Electrical centrifuge, which spins liquid samples at high speed.

Specifications

Some restricted information has been hidden. Sign in to see this information

Quality standards

  • EN IEC 61010-2-020, 2017, edition 3, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-020: Particular requirements for laboratory centrifuges
  • EN 12547, 2014, Centrifuges - Common safety requirements

Material

  • Metal housing
  • Stainless steel centrifuging chamber
  • Automatic lid locking
  • Imbalance switch-off
  • Automatic rotor recognition
  • Motor overheating protection

Components

Desktop centrifuge ROTINA 420 without rotor (ref. 4701)

Technical specifications

  • Maximum speed: ± 15000 rpm
  • Maximum RCF (Relative Centrifugal Force): 16,504 for 94 ml tubes, up to 12,007 for 250 ml bottles, up to 24,400 for reaction tubes
  • Running time: 1 s - 999 min, infinite continuous run, short cycle mode
  • Digital display
  • Power supply: 200-240 V, 50 Hz
  • Consumption 1500 VA
  • Lid locking

Dimensions

  • H x W x D: 423 x 506 x650 mm
  • Weight: approx. 75 kg

To be Ordered Separately

  • Swing-out rotor 4-places without bucket ref. 4784-A
  • 4x bucket/carrier for rotor (4784-A) ref. 4780
  • 4x Watertight cover for bucket (4780) ref. 4783
  • 4x Adaptor for 5 x 50 ml conical tube Falcon type ref. 4770
  • 4x Adaptor for 13 x 15 ml tubes ref. 4769

Instructions for use

Read carefully the supplier “Instruction Manual” before first use.

Please consult the “Updated laboratory procedures, 2022” available online via the Laboratory working Group sharepoint page: Laboratory Procedures and Resources.

https://msfintl.sharepoint.com/sites/msfintlcommunities/LabWG/SitePages/Laboratory-Manual-page.aspx

For offline access, contact your laboratory advisor.

Some restricted information has been hidden. Sign in to see this information

Precautions for Use

Secure all biomedical electrical devices in the laboratory. Contact your biomedical referent to define the best option.

Failing that, use a UPS double conversion for the electrical protection of the apparatus (see related articles below).

MSF requirements

Centrifuge with closed buckets to sediment mycobacteria from biological specimens, for inoculation on different media.