FACE MASK NIV, non-vented, complete, reusable, L
Valid Article
NIV FACE MASK, non-vented, reusable
Definition
A non-sterile, flexible, form-shaped device designed to be placed over a user's nose and mouth to interface with a continuous positive airway pressure (CPAP) or bi-level positive airway pressure (BPAP) unit to provide the respiratory tract with direct ambient air, or medical oxygen (O2) and air, at a higher pressure than ambient air for noninvasive positive pressure ventilation (NPPV).
It is typically made of plastic/silicone materials that create an airtight seal around the nose and mouth. It is available in various designs and sizes and stabilized with a headstrap.
Specifications
(Respireo non vented mask)
- a) mouth-and-nose cushion (rim)
- b) main body
- c) forehead support
- d) forehead rest
- e) adjusting knob
- f) adjusting screw
- g) cap
- h) elbow
- i) headgear
- l) clip
Technical specifications
- Non-vented mask, covers nose and mouth
- Standard female 22 mm tapered fitting for external connection to the ventilator
- Rotating elbow, coloured blue to indicate the mask is non-vented for dual limb circuit
- Physical dead space (= empty volume of the mask to the end of the swivel) : < 250 ml for the Large size
- Therapy pressure 4 to 40 cm H2O
- Does not contain latex, PVC or DEHP materials
- Reusable: the mask can be autoclaved at 120°C
Packaging & Labelling
Unit packaging including instructions for use.
To be Ordered Separately
For the Respireo Hospital F mask: the headgear has to be replaced every 7 days and is single patient use. Spare headgears have to be ordered separately. (see accessories)
Instructions for use
Precautions for Use
- This mask is not for use on patients with impaired laryngeal reflexes or other conditions predisposing to aspiration in the event of regurgitation or vomiting.
- The mask should not be worn unless the ventilator system is turned on and operating properly.
- Using a mask may cause tooth, gum or jaw soreness or aggravate an existing dental condition.
- Ensure the soft sleeves are fitted at all times during use.
Maintenance
Must be pre-disinfected, cleaned and sterilized by steam autoclave after each use.
Follow the manufacturer's instruction
(Cf Introduction: Disinfection and sterilization in the field)
MSF requirements
Volume or pressure cycled ventilators such as the Monnal T60 or LTV 1200 are designed to function with a closed circuit (non-vented) and the exhalation valve on the machine releases the expired/exhaled gases, while the inspiration valve is closed.
Therefore if there is a leak in the system such as with a vented mask, the inspiration valve will not close and an alarm will trigger continuously.