(MC2/SEAL) FORCEPS BIPOLAR straight 20cm,8x1mm,reus.VPB20D10

STD EEMDESUA1004

Valid Article

CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
Z12010980 - Electrosurgery instruments – hardware accessories
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

See Related

Add article to list

   

Add Kit To Wishlist

   

(ESU) ACTIVE BIPOLAR ELECTRODE

Definition

The active electrode delivers the RF current from the generator to the surgical site. The current to tissue action at the tip of the active electrode is high electron flow over a small area. As resistance is encountered heat is produced. The current density differs for each tip based on the size and shape of the tip.

With the bipolar electrosurgical method a bipolar device, often a forceps, is used. The electrical current passes from one side of the forceps, through the target tissue to the other side of the forceps, then back to the generator. The electrical current is restricted to the tissue between the tips of the forceps, therefore not requiring the use of a return electrode

Image

Specifications

Some restricted information has been hidden. Sign in to see this information

Technical specifications

  • ​bipolar forceps model bayonet cushing
  • total length: 18 or 20 cm
  • point: 2.0 mm

To be Ordered Separately

SSCOETCP55S: CLEANING PAD for electrosurgical tip, 5x5cm, sterile

Active electrode tips made from stainless steel have the disadvantage of eschar build up which results in increased resistance. This resistance interferes with the desired tissue effect and can become a fire hazard. These tips have to be cleaned with a scratch pad.

Instructions for use

Precautions for Use

  • Active HF instruments and HF handles must not be placed on the patient in order to avoid any injuries in case of involuntary HF activation
  • For cleaning the active electrode, you must disconnect the instrument from the generator. Directly after activation the tips could be hot. This may lead to burns.

Maintenance

The reusable electrode-handpieces and the electrodes must be cleaned and sterilized by steam autoclave after each use. Always leave the electrodes to cool down after sterilization.

(Cf Introduction: Disinfection and sterilization in the field)

Some restricted information has been hidden. Sign in to see this information