(densitometer Biosan) CALIBRATION KIT SD2350
Definition
The calibration kit is designed to test densitometers from Biosan.
Regular use of the kit allows to follow the intrument's precision, and to monitor the quality of the inoculum.
Specifications
Some restricted information has been hidden. Sign in
to see this information
Components
Latex particles suspended in a buffer solution
Technical specifications
- Do not open the tubes
- Do not shake the tubes
- Opaque liquid
Packaging & Labelling
The kit contains 5 McFarland Standard (0.5, 1.0, 2.0, 3.0 and 4.0), une density card and a package insert
Instructions for use
- It is recommended to test the accuracy of the densitometers once a month.
- Before testing the instrument, it must be cleaned.
- Each tube should be wiped clean with a dry, lint-free cloth prior to use.
Some restricted information has been hidden. Sign in
to see this information
Storage
- Store between 15°C and 30°C.
- Do not freeze.
- Do not use after expiry date.
- Shelf life: 9 months.
- Guaranteed minimum remaining shelf life at delivery: 1/3 of the total shelf life.
MSF requirements
Reserved for bacteriology programmes.