(syr. pump Agilia SP) MAINTENANCE TOOLS Z178957
STD
EIMTSYPE401
Valid Article
HS Code:
903300
Last Updated on:
07/11/2024, 19:01:26
Former
Code(s):
EEMDSYPS410
(syringe pump Agilia SP) MAINTENANCE TOOLS
Definition
This kit used to perform the preventive maintenance and the calibration of the syringe pump Agilia SP range.
Specifications
Some restricted information has been hidden. Sign in
to see this information
Technical specifications
Maintenance kit
- To calibrate the syringe barrel clasp
- C1 diameter calibration gauge (1)
- C2 diameter calibration gauge (2)
- C3 diameter calibration gauge (3)
- To calibrate the plunger displacement
- 20 mm displacement calibration gauge (4)
- 115 mm displacement calibration gauge (5)
- To calibrate and control the force sensor
- Dynamometer with arm spacer for calibration (6)
- To control the syringe barrel clasp
- V1 diameter verification gauge (7)
- V2 diameter verification gauge (8)
- V4 diameter verification gauge (9)
- To control the plunger displacement
- 65 mm displacement verification gauge (10)
- Screwdrivers
- Torx T10 screwdriver (11)
- Torx T8 screwdriver (12)
Agilia USB cable
To be Ordered Separately
- Agilia Partner Software
- USB cable Z073491 EEMDINPS406
Instructions for use
Kit for performing preventive maintenance and calibration of the Agilia SP range syringe pump
Precautions for Use
It is necessary to have training to use this kit...
Maintenance
As per Manufacturer recommendation the maintenance kit has to be checked by an accredited body :
- Every 12 months for all gauges and every 6 months for dynamometer or
- If the tool fell or could be damaged
MSF requirements
Maintenance kit compatible with the Agilia SP
Some restricted information has been hidden. Sign in
to see this information