REFRIGERATOR (Vestfrost R-427i) 361l net, medic., glass door
NST
ELAEREFE02-
Valid Article
HS Code:
901890
Last Updated on:
07/11/2024, 18:56:02
Hazardous materials and products (identified by their 4-digit transport UN-number) must be transported under certain conditions as specified in the MSDS sheet (Material Safety Data Sheet).
REFRIGERATOR (Vestfrost R-427i) 361l net, glass door
Definition
A mains electricity (AC-powered) fridge designed for the storage of pharmaceutical or medical products at refrigerated temperatures in pharmacies, laboratories and hospitals. It is provided with a built-in alarm system that warns against unexpected temperature variations.
Specifications
Some restricted information has been hidden. Sign in
to see this information
Quality standards
DIN 13277, 2022, Refrigerators and freezers for laboratory and medical applications - Terminology, requirements, testing
Technical specifications
- Audible and visual alarm
- High/Low Temp. Alarm
- Open Door Alarm
- Probe Failure Alarm
- Power Failure Alarm
- Top LED light provides maximum visibility
- Reversible glass door: Closure < 90°, Hold > 90°
- Integrated lock
- Adjustable wire shelve
- Net volume: 361 litres
- Temperature range: +2 to +20C°
- Voltage 220V, Frequence 50-60Hz
- Energy consumption: 0.4kWh/24h
- Battery backup: 48h
- Insulation tickness 50mm
- Insulation type: Polyurethane with cyclopentane
- Refrigerant: type R600a / 60 g
- Shelves
- No: 5
- material shelf: perforated aluminium
- Max Loading per Shelf, 20kg
- Operating conditions
- temperature max 32°C
- relative humidity max 65%
Dimensions
- Dimensions (HxWxD):
- outer 2002x595x642cm
- inner 1827x475x495
- Weight:
- gross: 88kg
- net: 74 kg
- Volume
- gross 427 l
- net: 361 l
MSF requirements
Specific large refrigerator with glass door for medical products, intended for laboratories with high activity. It is used for products requiring cold chain storage like rapid diagnostic tests, laboratory analyzers' tests, reagents...
Some restricted information has been hidden. Sign in
to see this information
Some restricted information has been hidden. Sign in
to see this information
Description approved by referent of OCB (Jonathan Delchambre) in January 2024.