(MicroMGIT) FLUORESCENCE READER [BD-445923]
STD
ELAEFLUE3--
Valid Article
Account code:
60110
HS Code:
901890
Last Updated on:
22/11/2024, 22:17:57
Former
Code(s):
-X ELAEZBD0206 ELAEFLUO300
W02030299 - Instruments for blood cultures and mycobacteria - other
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
FLUORESCENCE READER (MicroMGIT)
Reserved for TB culture programmes.
Definition
Handheld battery-powered instrument with positive/negative readout for fluorescence detection of MGIT mycobacteria growth indicator tube 4 ml.
Specifications
Some restricted information has been hidden. Sign in
to see this information
Technical specifications
- Power: 9 Volt rectangular alkaline battery
- Operating temperature: +15°C to +40°C
- 0 to 90% relative humidity, non-condensing
- No direct sunlight and no direct heat
- Supplied with alkaline battery and calibration tube
Dimensions
- H x W x D : 12.7 x 8.9 x 15.2 cm
- Weight: 333 g
Instructions for use
Calibrate the instrument with calibration tube before each use.
Sample readings must be interpreted only after positive and negative control reading.
Confirm any positive result with AFB microscopy.
Precautions for Use
Secure the electrical protection of the apparatus with a double conversion UPS (see related articles below).
Storage
Temperature storage: 0 to 45°C
MSF requirements
Manual reader used to detect microorganism growth in MGIT test tubes.
Some restricted information has been hidden. Sign in
to see this information
Some restricted information has been hidden. Sign in
to see this information