(MicroMGIT) FLUORESCENCE READER [BD-445923]

STD ELAEFLUE3--

Valid Article

Former Code(s): -X ELAEZBD0206 ELAEFLUO300
CE marking: declaration that the product meets EU standards for health, safety, and environmental protection. The CE marking indicates that the product may be sold freely in any part of the European Economic Area, regardless of its country of origin.
W02030299 - Instruments for blood cultures and mycobacteria - other
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
The order of this product needs to be justified and is only acceptable under certain conditions.
OC subscriptions: included in supply or field order tools (e.g. UniField)
SC subscriptions: included in supply or field tools
OC validations: approved for procurement and use by an OC for international or local orders specifying context and activity in Medical Standard Lists (MSL)

This combination does not exist.

Documents See Related

Add article to list

   

Add Kit To Wishlist

   

FLUORESCENCE READER (MicroMGIT)

Reserved for TB culture programmes.

Definition

Handheld battery-powered instrument with positive/negative readout for fluorescence detection of MGIT mycobacteria growth indicator tube 4 ml.

Specifications

Some restricted information has been hidden. Sign in to see this information

Technical specifications

  • Power: 9 Volt rectangular alkaline battery
  • Operating temperature: +15°C to +40°C
  • 0 to 90% relative humidity, non-condensing
  • No direct sunlight and no direct heat
  • Supplied with alkaline battery and calibration tube

Dimensions

  • H x W x D : 12.7 x 8.9 x 15.2 cm
  • Weight: 333 g

Instructions for use

Calibrate the instrument with calibration tube before each use.

Sample readings must be interpreted only after positive and negative control reading.

Confirm any positive result with AFB microscopy.

Precautions for Use

Secure the electrical protection of the apparatus with a double conversion UPS (see related articles below).

Storage

Temperature storage: 0 to 45°C

MSF requirements

Manual reader used to detect microorganism growth in MGIT test tubes.

Some restricted information has been hidden. Sign in to see this information
Some restricted information has been hidden. Sign in to see this information