(spectrophotometer) KIT,BILIRUBIN direct/total 2x100ml 10740
Valid Article
(spectrophotometer) KIT, BILIRUBIN D&T
Definition
Kit of reagents to determine the dosage of bilirubin in a blood sample (non hémolysed serum or heparinised plasma) with the use of a spectrophotometer, following the Jendrassik/Grof method.
To be used with the Humalyzer 2000 which allows testing of the main clincal chemistry and haematology tests, including bilirubine.
It replaces the Bilirubin kit (spectrophotometer), ref. 10012 (ELASPET10012).
Specifications
Components
Contents of kit:
- 1 x 100 ml Total bilirubin reagent (white cap)
- Sulphanilic acid 14 mmol/l
- Hydrochloric acid 300 mmol/l
- Caffeine (accelerator) 200 mmol/l
- Sodium benzoate 420 mmol/l
- 1x 9 ml T-Nitrite reagent (white cap)
- for determination of total bilirubin
- Sodium nitrite 390 mmol/l
- 1 x 100 ml Direct bilirubin reagent (blue cap)
- Sulphanilic acid 14 mmol/l
- Hydrochloric acid 300 mmol/l
- 1 x 9 ml D-nitrite reagent (blue cap)
- for determination of direct bilirubin
- Sodium nitrite 25 mmol/l
Note: Both reagents and nitrite solutions are ready for use.
Technical specifications
- Complete kit for 200 measurements (controls, standard, samples): 100 total bilirubin and 100 direct bilirubine
- Colorimetric test, Jendrassik/Gróf method (546 nm)
- Sample types: serum, heparinised plasma
- Wavelength: 546 nm
- Operating temperature: 20 to 25°C
- Optical path: 1 cm
- Total bilirubin reagent: yellowish fluid
- T-nitrite, Direct bilirubin and D-nitrite reagents: colourless fluid
Packaging & Labelling
Kit with 2 vials of 100 ml, and 2 vials of 9 ml
Transport Dangerous Goods
Transport not regulated
To be Ordered Separately
- Lyophilised control Serodos (Human): normal range, ref. 13951
- Lyophilis control Serodos Plus (Human): abnormal range, ref. 13151
(see related articles below)
Instructions for use
Follow the instructions in the package insert.
Please consult the “Updated laboratory procedures, 2022” available online via the Laboratory working Group sharepoint page: Laboratory Procedures and Resources.
https://msfintl.sharepoint.com/sites/msfintlcommunities/LabWG/SitePages/Laboratory-Manual-page.aspx
For offline access, contact your laboratory advisor.
Storage
- Do not freeze
- Shelf life: 8 months
- Not to be used beyond expiry date
- Reagents are stable after opening if stored between 15 and 25°C
Classification EC Regulation N° 1272/2008
Not considered as hazardous
Direct bilirubin reagent : contains sulphanilic acid and may produce allergic reaction (EUH208)
P401 | Store in accordance with local and regional/ national/ international regulations |
P234 | Keep only in original container. |
P260 | Do not breathe dust/fume/gas/mist/vapours/spray. |
P262 | Do not get in eyes, on skin, or on clothing. |
P281 | Use personal protective equipment as required. |
P303 + P361 + P353 | IF ON SKIN (or hair): Remove/Take off immediately all contaminated clothing. |
P305 + P351 + P338 | IF IN EYES: rinse cautiously with water for several minutes. If the victim wears contact lenses, remove them if possible. Keep rinsing |
P337 + P313 | If eye irritation persists: get medical advice/attention |
P501 | Dispose of contents/container to location in accordance with local and regional/ national/ international regulations |