(pump Flocare) ADMINISTRATION SET, ENFit, ster.,s.u. 586514
STD
EEMDENPC102
Valid Article
Account code:
60210
HS Code:
901890
Last Updated on:
07/11/2024, 18:53:22
Single use
A single-use device, also referred to as a disposable device, is intended for use on one patient during a single procedure. It is not intended to be reprocessed (i.e., cleaned and disinfected or sterilized).
Z12030385 - Infusion instruments – consumables
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
In Europe, medical material that fulfills the definition of a medical device according to the MDR (Medical Device Regulation) is classified into 4 classes
(pump Flocare) ADMINISTRATION SET
Two models of administration sets can be ordered according to the brand of your feeding pump.
The Applix pump is no longer produced but the sets can still be ordered.
Definition
A device intended to enable connection of components of an enteral feeding system using an electrical pump. It is typically a small plastic device containing a lumen to facilitate enteral administration of feeds after connection; it has no additional functionality and is dedicated to enteral feeding applications. This is a single-use device.
Specifications
Some restricted information has been hidden. Sign in
to see this information
Quality standards
- ISO 80369-1, 2018, edition 2, Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
- ISO 80369-3, 2016, edition 1, +A1 2019 Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications
Technical specifications
- TUBING:
- Transparent PVC, DEHP free, latex free
- Ø internal = 2.7 mm
- Length: 175 – 200 cm
- DROP-COUNTING CHAMBER
- THUMB WHEEL REGULATOR: Allows to adjust feeding flow rate between 0 and the maximal value. Alternative: CLAMP for positioning the pump
- PROXIMAL CONNECTOR (to the feeding pouch): ENPlus, non compatible with Luer
- ADMINISTRATION PORT: For medication, hydration and flushing of the tubing, located at 40 cm from the end
- DISTAL CONNECTOR (towards the nasoenteral tube): connection ENFit, non compatible with luer
- COLOUR of the ENFit connections: purple is often used for enteral feeding medical devices, but the specific ISO norm does not specify the colour.
- Sterile
- For single use
Packaging & Labelling
Unit sterile packaging in peel-open pack
Instructions for use
- Connect the set to the feed container as instructed on the packaging of the pump set.
- Tube feeding may only commence after correct position of the tube has been validated (e.g. by pH measurement of gastric contents)
- Follow aseptic practice when handling feed, tubes and feeding sets
- Flush the tube before and after feed and medication administration, and at least 3 times a day
- Sets need to be changed every 24 hours
Precautions for Use
Preventing misconnection errors
- Regularly inspect tubing for proper connections.
- Reconnect tubing only under good lighting. Trace tubing back to the source prior to reconnecting.
- Do not modify or try to adapt connections to enteral devices.
- Do not use IV pumps to administer enteral feedings.
- Use enteral syringes (that cannot connect with an IV line) to deliver medications enterally.
MSF requirements
Administration set that is required to administer enteral feeding with the enteral nutrition/feeding pump.