HAEMATOLOGY ANALYSER (Sysmex XQ320), 110-240V, 50-60Hz
Valid Article
HAEMATOLOGY ANALYSER (Sysmex XQ-320)
The XQ-320 replaces the XP-300 starting from 2023. The XQ-320 uses the same reagents and solutions as the XP-300 but spare parts are different for both models.
Definition
Sysmex XQ-320 is an automated haematology analyser allowing analysis of the main clinical parameters associated to haematology, including, total and differential (3-part) white blood count, platelets and haemoglobin.
Specifications
Parameters: haemoglobin measurements, white cell, red cell and platelet counts, red cell indices: 20 parameters
| - WBC | - NEUT% | 
| - RBC | - LYM# | 
| - HGB | - MXD# | 
| - HCT | - NEUT# | 
| - MCV | - RDW-SD | 
| - MCH | - RDW-CV | 
| - MCHC | - PDW | 
| - PLT | - MPV | 
| - LYM% | - P-LCR | 
| - MXD% | - PCT | 
Quality standards
- EN IEC 61010-1, 2010, edition 3, +2019 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
 - EN IEC 61010-2-081, 2020, edition 3, Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes
 - EN IEC 61010-2-101, 2022, edition 3, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
 
Components
Haematology ananlyser (apparatus)
- open cap analyser incl. 2D barcode reader
 - Power cable, IFU and CELLPACK spout kit (83400116)
 - Wireless Keyboard (international Layout)
 - Single sheet certificate
 - PRE-INSTALLATION SERVICE (which is not a product)
 
Technical specifications
- Standalone device
 - All parameters using the same reagents (see related articles below)
 - Measurement principle: Impedance measurement (WBC, RBC/PLT)
non-cyanide haemoglobin detection method (HGB) - Analysis mode: whole blood (WB) mode and pre-dilution (PD) mode
 - Aspiration volume: 16 μL in whole blood mode, 65 μL in pre-diluted mode
 - Throughput: approx. 70 tests/h (WB mode), 60 tests/h (PD mode)
 - Time to result: <1 minute
 - Operating temperature: 10 to 40 °C
 - Operating humidity: 20-85%
 - Fully automated and standardised measurement
 - Automated background check analysis to avoid carryover
 - Auto dispense function for standardised PD measurement
 - Interface: RS-232C, VGA, 2x Ethernet (RJ45), 4x USB
 - Operating system: Embedded Microsoft ® Windows 10 with McAfee Security
 - Power supply: AC 100 to 240 V (50/60 Hz)
 
Dimensions
Dimensions: 365 x 440 x 450 mm
Weight: 22 kg
Transport Dangerous Goods
- UN3091: Lithium metal batteries contained in equipment
 - Class: 9
 - Packing instructions: 970 Section II
 - Limit per package:
 - Pax A/C (Passengers & Cargo Aircraft) = 5 kg
 - CAO (Cargo Aircraft Only) = 5 kg
 - Not regulated for transport by sea
 
Supplied with the Article
- CELLPACK 10L
 - CELLCLEAN
 - STROMATOLYSER-WH 3X500ML
 
Instructions for use
Please consult the “Updated laboratory procedures, 2022” available online via the Laboratory working Group sharepoint page: Laboratory Procedures and Resources.
https://msfintl.sharepoint.com/sites/msfintlcommunities/LabWG/SitePages/Laboratory-Manual-page.aspx
For offline access, contact your laboratory advisor.
Precautions for Use
- Air conditioning is necessary. Dust-free environment.
 - Installation to be done by authorised Sysmex distributor.
 - Secure all electrical devices in the laboratory with a UPS double conversion (see related articles below). To define the best sizing of the UPS, contact your biomedical or energy referent.
 
Maintenance
A maintenance contract is mandatory. Six-monthly maintenance to be performed by authorised Sysmex distributor.