AUTOCLAVE 28L (Tuttnauer 2840EL-D) laboratory benchtop, 230V
STD
ELAEAUTE02-
Valid Article
Account code:
60110
HS Code:
841920
Last Updated on:
07/11/2024, 17:48:07
Z120113 - Cleaning, disinfection and sterilisation instruments
European Medical Device Nomenclature (EMDN) is the nomenclature of use by manufacturers when registering their medical devices in the EUDAMED database. EMDN is characterised by its alphanumeric structure that is established in a seven-level hierarchical tree.
AUTOCLAVE 28L (Tuttnauer 2840EL-D) laboratory benchtop
Definition
Benchtop steam autoclave for laboratory use, for the sterilisation of the media, material and destruction of biohazard waste.
Specifications
- Chamber volume: 28.5 liters.
- Loading capacities:
- 9 erlenmeyers of 250 ml or 4 erlenmeyers of 500 ml or 1 erlenmeyer of 1000 ml
- 12 Schott-Duran flasks of 250 ml or 9 Schott-Duran flasks of 500 ml ou 5 Schott-Duran flasks of 1000 ml
- Sterilisation temperature range: 105 to 138°C
Some restricted information has been hidden. Sign in
to see this information
Quality Standards Comment
- EN 13060:2014+A1 : 2018: Small steam sterilizers
- IEC 61010-2-040 : 2020: Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-040 Particular requirements for sterilizers and washer-disinfectors used to treat medical materials
- IEC 61326-1: 2020 EMC : Electrical equipment for measurement, control and laboratory use - Part 1: general requirement
Components
- Chamber constructed of 316L grade stainless steel.
- Door, generator and trays are made from stainless steel.
- Water tank is in hard plastic.
- Door handle is made of hard plastic,thermo-isolated and can be touched without risks.
Technical specifications
- Use conditions:
- To be used inside
- Maximum external temperature for use: 40°C
- Maximum relative humidity for use: 85%
- The altitude should be maximum 2000 m
- Electrical informations:
- 1Ph, 230V/50Hz, 10A.
- It is recommended that a ground fault protection device be installed
- Other features:
- Front loading
- The temperature and pressure are controlled through sensors placed inside and outside the media container or bottles. PID (proportional integrak differential) pressure control
- 8 factory set cycle programs
- Multi-color control panel
- 26 languages
Dimensions
- Chamber dimensions: 280 (Ø) x400 mm
- External dimensions (WxHxD): 530 x 440 x 630 mm (length with door open: 961 mm)
- Net weight: 43 kg
- Shipping dimensions (WxHxD): 620 x 510 x 720 mm
- Shipping weight: 51 kg
To be Ordered Separately
Check with your biomedical referent what spare parts and accessories should be ordered with the autoclave
Instructions for use
Some restricted information has been hidden. Sign in
to see this information
MSF requirements
Specific for microbiology/culture laboratory, after validation by the advisor.
Some restricted information has been hidden. Sign in
to see this information